SMWA12C Datasheet, PDF (1/2 Pages) Vitec Power – 12W Medical Grade Wall Mount Power Supply
SMWA12C SERIES
12W Medical Grade Wall Mount Power Supply
– Wide Input Voltage 90 to 264 VAC, 47 to 63Hz
– European type 2 prong plug
– Single Output
– Output Voltage Available From 5VDC Thru 27VDC
– Over Voltage, Over Load, and Over Temperature Protection
– Class II Insulation
– Energy Star 2.0, CEC Level V, and RoHS Compliance
3 Year Warranty
Approvals:
Single Output
Part Number
Output Voltage
Max. Output Current
SMWA12C-S02
5 ~ 6 VDC
2.00 ~ 1.66 A
SMWA12C-S03
6 ~ 8 VDC
2.00 ~ 1.50 A
SMWA12C-S04
8 ~ 11 VDC
1.50 ~ 1.09 A
SMWA12C-S05
11 ~ 13 VDC
1.09 ~ 0.92 A
SMWA12C-S06
13 ~ 16 VDC
0.92 ~ 0.75 A
SMWA12C-S07
16 ~ 21 VDC
0.75 ~ 0.57 A
SMWA12C-S08
21 ~ 27 VDC
0.57 ~ 0.44 A
The total regulation on model S02~S03 is required to use AWG#18 / 6FT output cable.
The total regulation on model S04~S08 is required to use AWG#20 / 6FT output cable.
The regulation and efficiency are not guaranteed if changes the output cable
Total Regulation
5%
5%
5%
5%
5%
3%
3%
Electrical Characteristics
Parameter
Test Conditions
Min.
Typ.
Input Voltage
Input Frequency
Output Power Range
Input Current (Low Line)
Input Current (High Line)
Low Line Inrush Current
High Line Inrush Current
Efficiency
Line Regulation
Load Regulation
Over Voltage Protection
Over Current Protection
Transient Response
Hold-Up Time
Start Up Time
Ripple & Noise(Peak to Peak)
Safety Ground Leakage Current
Temperature Coefficient
No-Load Power Consumption
Operating Voltage
Vin=90 to 264VAC
Io=Full load, Vin=115VAC
Io=Full load, Vin=230VAC
Io=Full load, 25°C Cool start, Vin=115VAC
Io=Full load, 25°C Cool start, Vin=230VAC
Io=Full Load, Vin=230VAC
Io=Full Load
Vin=230VAC
Io=Full Load to Half Load, Vin=100VAC
Io=Full Load, Vin=110VAC
Io=Full Load, Vin=100VAC
Full Load, Vin=90VAC
Io=Full Load, Vin=240VAC
All output
No load, Vin=240VAC
90
47
0
0.25
0.17
14
28
73.3
77.7
0.5
3
112
110
10
16
0.3
1
0.5
-0.04
Thermal Shutdown By Junction
Temperature Controller *
The parameter is not subject to production
test-verified by design/characterization of integrated
-25
controller. Auto recovery.
*As long as faulty conditions have been removed , the adaptor will automatically power up as usual.
Conditions
Parameter
Test Conditions
Operating Temperature
Storage Temperature
Relative Humidity
Operation temperature at 25°C, calculated per MIL-HDBK-217F
Derate linearly from 100% load at 50°C to 50% load at 70°C
Min.
Typ.
0
50
-40
5
0.1M
Max. Output Power
10W
12W
12W
12W
12W
12W
12W
Max.
264
63
12
0.35
0.22
16
35
85
1
5
132
150
4
2
1
0.1
0.04
0.3
130
Unit
VAC
Hz
W
A
A
A
A
%
%
%
%
%
mS
mS
S
%
mA
%/°C
W
°C
Max.
Unit
70
°C
85
°C
95
%