SMDA130 Datasheet, PDF (1/2 Pages) Vitec Power – 130W Desktop Power Supply for Medical Equipment
SMDA130 SERIES
130W Desktop Power Supply for Medical Equipment
− Wide Input Voltage 90 to 264 VAC, 47 to 63Hz
− IEC-320-C14 input inlet
− Input Surge Current, Over Voltage, Over Load and Output Voltage Protection.
− Class I Insulation
− Active Power Factor Correction
− Energy Star 2.0, CEC V, and RoHS compliance
− ON/OFF Switch (Optional)
3 Year Warranty
Approvals:
Product Number
Output Voltage
SMDA130-S05
SMDA130-S08
12 VDC
24 VDC
Total Regulation is conditioned by below configuration
(S05: AWG16/2C/4FT output cable)
(S08: AWG18/2C/6FT output cable)
Single Output
Max. Output Current
10.84 A
5.42 A
Total Regulation
5%
3%
Parameter
Input Voltage
Input Frequency
Power Factor Correction
Output Power Range
Input Current (Low Line)
Input Current (High Line)
Low Line Inrush Current
High Line Inrush Current
Efficiency
Line Regulation
Load Regulation
Over Voltage Protection
Over Current Protection
Transient Response
Hold-Up Time
Start Up Time
Ripple & Noise (Peak to Peak)
Safety Ground Leakage Current
Temperature Coefficient
Electrical Characteristics
Test Conditions
Operating Voltage
Io=Full load, Vin=230 VAC
Vin= 90 to 264 VA C
Io=Full load, Vin=115 VAC
Io=Full load, Vin= 230 VAC
Io=Full load, 25°C, Cool start,
Vin=115VAC
Io=Full load, 25°C, Cool start,
Vin=230VAC
Io=Full Load, Vin=230VAC
Io=Full Load
Vin=230VAC
Io=Full Load to Half Load, Vin=100VAC
Io=Full Load, Vin=110VAC
Io=Full Load, Vin=100VAC
Full Load, Vin=90VAC
Io= Full Load, Vin=240VAC/60Hz
All output
Min.
Typ.
90
47
0.95
0
88
0.5
5
112
110
16
-0.04
Maximum
Output Power
130W
130W
Max.
264
63
1
130
1.32
0.66
30
50
1
132
150
4
2
1
0.1
0.04
Unit
VAC
Hz
W
A
A
A
A
%
%
%
%
%
mS
mS
S
%
mA
%/°C