English
Language : 

SHFA60 Datasheet, PDF (1/2 Pages) Vitec Power – 60W Open-Frame Power Supply for Medical Equipment
SHFA60 SERIES
2 Year Warranty
60W Open-Frame Power Supply for Medical Equipment
 Wide Input Voltage 90 to 264 VAC, 47 to 63Hz
 Output Voltage Available From 12VDC Thru 40VDC
 Input connector mates with molex housing 09-50-3031/35977-0300 and molex 2478/
35922 series crimp terminal.
 Output connector mates with molex housing 09-50-3061/35977-0600 and molex
2478/ 35922 series crimp terminal.
 Single Output
 Size: 2"x4"x1"
 Class I Insulation
 Input Surge Current, Over Load and Over Voltage Protection
 Solution for Health Care Equipment
Approvals:
Product Number
SHFA60-S05
SHFA60-S06
SHFA60-S07
SHFA60-S08
SHFA60-S09
SHFA60-S10
Output Voltage
12 ~ 13 VDC
13 ~ 16 VDC
16 ~ 21 VDC
21 ~ 27 VDC
27 ~ 33 VDC
33 ~ 40 VDC
Single Output
Output Current
5.00 ~ 4.61 A
4.61 ~ 3.75 A
3.75 ~ 2.85 A
2.85 ~ 2.22 A
2.22 ~ 1.81 A
1.81 ~ 1.50 A
Total Regulation
5%
5%
5%
3%
3%
3%
Parameter
Input Voltage
Input Frequency
Output Power Range
Input Current (Low Line)
Input Current (High Line)
Low Line Inrush Current
High Line Inrush Current
Efficiency
Line Regulation
Load Regulation
Over Voltage Protection
Over Current Protection
Transient Response
Hold-Up Time
Start Up Time
Ripple & Noise (Peak to Peak)
Safety Ground Leakage Current
Temperature Coefficient
Electrical Characteristics
Test Conditions
Operating Voltage
Vin= 90 to 264 VAC
Io=Full load, Vin=115 VAC
Io=Full load, Vin= 230 VAC
Io=Full load, 25°C, Cool start,
Vin=115VAC
Io=Full load, 25°C, Cool start,
Vin=230 VAC
Io=Full Load, Vin=230VAC
Io=Full Load
Vin=230VAC
Io=Full Load to Half Load, Vin=100VAC
Io=Full Load, Vin=110VAC
Io=Full Load,
Vin=100VAC
Full Load, Vin=90VAC
Io= Full Load, Vin=240VAC
All output
Min.
90
47
0
0.3
112
110
12
0.3
-0.04
Maximum Output Power
60W
60W
60W
60W
60W
60W
Typ.
26
43
85
16
1
0.1
Max.
264
63
60
1.62
0.72
30
47
1
5
132
150
4
2
1
0.3
0.04
Unit
VAC
Hz
W
A
A
A
A
%
%
%
%
%
mS
mS
S
%
mA
%/°C