English
Language : 

MUU61 Datasheet, PDF (1/2 Pages) Volgen Kaga Electronics Inc. – 63W U-bracket Switching Power Supplies For Medical Equipment.
MUU61 SERIES
63W U-bracket Switching Power Supplies For Medical Equipment.
Description:
The MUU61 series of compact, open frame constructed, AC/DC
switching mode power supplies provide 63 Watts of continuous
output power . They are ideally suited for use in hospital instrument
and many other applications. All models meet FCC Part-18 class B
and CISPR-11 EN55011 class B emission Limits and are designed
to comply with UL/c-UL(UL 60601-1) ,TUV/T-mark(EN 60601-1)
and new CE requirements. All units are 100% burned in and tested.
Features:
Wide Input Voltage 90 to 264 VAC,47 to 63 Hz
Internal EMI filter
Single to Quad Output
Input connector mates with Molex housing 09-50-3031 and
Molex 2478 series crimp terminal
Output connector mates with Molex housing 09-50-3081
and Molex 2478 series crimp terminal
Output Voltage Available From 3 VDC Thru 40 VDC
Input Surge Current, Over Voltage and Over Load protection
Power Fail Detect (Optional)
Class I Insulation
Two Years Warranty
Safety Approvals :
Electrical Characteristics:
Sym.
Parameter
Test Conditions
Min. Typ. Max. Unit
Vin Input Voltage
Operating Voltage
90
264 VAC
Fin Input Frequency
47
63 Hz
Wo Output Power Range
Vin=90 to 264 VAC
0
63
W
Vo Output Voltage Range
See rating Chart V
Io Output Current Range
See rating Chart A
Iil Input Current (Low Line) Io=Full load, Vin=115VAC
1.6 A
Iih Input Current (High Line) Io=Full load, Vin=230VAC
0.8 A
I rl Low Line Inrush Current
I rh
Eff
REG-i
R E G -o
High Line Inrush Current
Efficiency
Line Regulation
Load Regulation
Io=Full load, 25 C,Cool start,
Vin=115VAC
Io=Full load, 25 C,Cool start,
Vin=230VAC
Io=Full load, Vin=230VAC
Io=Full load
Vin=230VAC
15 18
A
21 25 A
70 80 85 %
0.5 1
%
36
%
OVP Over Voltage Protection
112
132 %
OCP Over Current Protection
110
150 %
Ttr Transient Response
Io=Full load to Half Load, Vin=100VAC
4
mS
Thold Hold-Up Time
Io=Full load, Vin=110VAC
16
mS
Ts Start Up Time
Io=Full load, Vin=100VAC
Vp-p Ripple & Noise (Peak to Peak) Full load, Vin=90VAC
Ilk Safety Ground Leakage Current Io=Full load, Vin=240VAC
0.3 1 2
S
12
%
0.1 0.3 mA
Tc Temperature Coefficient All output
-0.04
0.04 %/ C
Environmental :
Sym.
Parameter
Toper Operating Temperature
Te s t Conditions
Tstg Storage Temperature
Hr Relative Humidity
Pd Derate linearly from 100% load at 50 C to 50% load at 70 C
Min. Typ. Max.
0
70
-40
85
5
95
Unit
C
C
%
Safety Specifications:
Sym.
Parameter
Vps
Dielectric Withstanding Voltage
for Primary to secondary
Dielectric Withstanding Voltage
Vpg for Primary to Ground
Ri Isolation Resistance
CISPR EMI requirements for CISPR-11
FCC EMI requirements for FCC PART-18
Te s t Conditions
Primary to secondary
Primary to ground
Test Voltage=2100VDC
Vin=220VAC
Vin=110VAC
Min. Typ. Max. Unit
5600
VDC
2800
50
B
B
VDC
M
CLASS
CLASS