English
Language : 

TXM25_16 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – Medical Grade 25W Wall Mount Power Adapters
TRUMPower
FEATURES
• IEC 60601-1-11:2015 Home Healthcare Certification
• Approved to UL/IEC/EN 60601-1 3.1 Edition
• Meets IEC/EN 60601-1-2 4th Edition EMC
• Class II Construction with 2 MOPP
• Interchangeable AC Plugs
• UL, cUL, TUV, PSE, CCC, CB & CE Approvals
TXM25 Series
Medical Grade 25W Wall Mount Power Adapters
INPUT SPECIFICATIONS
Input Voltage Range ................. 80~275 VAC
Input Frequency ........................ 47~63 Hz
Input Current ............................. 0.7A rms @ 100 VAC,
0.4A rms @ 240 VAC
Inrush Current .......................... 50A @ 100 VAC or 100A
@ 240 VAC, at 25°C cold start
No Load Power Consumption... 0.3W max.
Enclosure Leakage Current ..... 100µA max. @ 240 VAC, 60 Hz
OUTPUT SPECIFICATIONS
Output Power Ratings .............. See models list
Minimum Load ......................... Not required
Total Regulation ....................... See models list
Ripple & Noise ......................... See models list
Overload Protection ................. Set at 110-150% of its nominal
output voltage; Auto recovery
Short Circuit Protection ............ Auto recovery
Temperature Coefficient .......... ±0.04%/°C max.
Transient Response ................ 4ms @ 110 VAC, full load
GENERAL SPECIFICATIONS
Efficiency .................................. See models list
Hold-up Time ............................ 12ms min. @ 100 VAC, full load
Start-up Time ............................ 3 seconds max. @ full load
Line Regulation ........................ ±1% max. @ full load
Operating Temperature ........... -10°C to +70°C
Derating .................................... Derate from 100% at +40ºC
linearly to 50% at +70°C
Storage Temperature ............... -40°C to +85°C
Relative Humidity ..................... 0% to 95% non-condensing
Withstand Voltage .................... 4,000 VAC from input to output
Operating Altitude ..................... 3,000 meters max.
MTBF ......................................... 100K hours minimum at full load,
25°C ambient, calculated per
MIL-HDBK-217F
STANDARDS & COMPLIANCES
IEC/EN 60601-1-2: 2014 .... EMC & Immunity Performance
EN 55011, CISPR11 ........... Class B, conducted & radiated
FCC, VCCI ........................... Class B, conducted & radiated
EN 61000-3-2 ...................... Harmonic distortion, Class A
EN 61000-3-3 ...................... Line fluctuations & flicker
EN 61000-4-2 ...................... ESD, ±15 KV air and ±8 KV contact
EN 61000-4-3 ...................... Radiated immunity, 10 V/m
EN 61000-4-4 ...................... Fast transient/burst, ±2 KV
EN 61000-4-5 ...................... Surge, ±1 KV diff., ±2 KV com.
EN 61000-4-6 ...................... Conducted immunity, 6 Vrms
EN 61000-4-8 ...................... Magnetic field immunity, 30 A/m
EN 61000-4-11 .................... Voltage dips,
30% reduction for 500ms,
60% reduction for 100ms,
>95% reduction for 10ms
Safety Standards ................. UL/IEC/EN 60601-1 (Edition 3.1),
............................................. ANSI/AAMI ES 60601-1 (2012),
............................................. CSA C22.2 No. 60601-1 (2014);
............................................. IEC 60601-1-11:2015
Agency Approvals ................ UL, cUL, TUV, PSE, CCC, CE, CB
Other Compliances ............ RoHS2, CEC
MECHANICAL SPECIFICATIONS
Unit: inch. [mm]
Tolerance: 0.02 [0.5] max.
Weight: 200g (0.44 lb) approx.
Notes:
1. Standard output connector: 2.5 x 5.5 mm or 2.1 x 5.5 mm barrel plug
2. Standard output cable: 4-ft, AWG#18 for 3-10.9V output models;
4-ft, AWG#20 for 11-32.9V output models; 4-ft, AWG#22 for 33-55V
output models.
3. Contact TRUMPower for other output cable and connector options.
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Building 13 • Santa Clara, California 95054, USA • 408-988-6616 • sales@trumpower.com • www.TRUMPower.com