English
Language : 

TWM65 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – 65W Medical & ITE Grade AC/DC Power Supplies
TRUMPower
TWM65 Series
65W Medical & ITE Grade AC/DC Power Supplies
FEATURES
• Convection-Cooled 65 Watts in 2" x 3" Footprint
• Medical & ITE Dual Approvals
• 2x MOPP Suited for BF (Body Floating) Applications
• Factory Preset 5V or 12.0V to 56.9V in 0.1V Increments Ouput
• Up to 0.95 Power Factor
• No Load Power Consumption < 0.2W
ELECTRICAL SPECIFICATIONS
Input Voltage ........................ 90-264 VAC
Input Frequency ................... 47-63 Hz
Input Current ........................ 1A (rms) max. @ 115 VAC,
0.6A (rms) max. @ 230 VAC
Inrush Current ..................... <65A max. @ 115 VAC, <130A max.
@ 230 VAC at cold start +25°C
Earth Leakage Current ........ 160µA max. @ 264 VAC
Power Factor ....................... >0.95 @115 VAC, >0.9 @230 VAC
Output Power Ratings ......... See table
No Load Power Cons .......... <0.1W on +5V output model;
<0.2W on other output models
STANDARDS & COMPLIANCES
EMC Standards ................... EN 60601-1-2, EN 55011/FCC Part
18 Class B, EN 55022/CISPR22/
FCC Part 15 Class B
Safety Standards ................. IEC/EN/UL 60601-1 3rd Ed.,
ANSI/AAMI ES 60601-1:2005,
CSA C22.2 No. 60601-1:08,
IEC/EN/UL 60950-1 2nd Ed.,
CSA-C22.2 No. 60950-1 2nd Ed.
Agency Approvals ................ UL, CUS, TUV, CE, CB, FCC
Other Compliance ............... RoHS
GENERAL SPECIFICATIONS
Efficiency .................................. >87% (except >85% for 5V)
at full load, 115 VAC
Switching Frequency ............... 65KHz
Hold-up Time ........................... >10 ms typical @ 115 VAC
Operating Temperature* .................. -20°C to +70°
Derating ................................... Derate linearly 2.5%/°C from
.............................................+41°C to +70°C
Storage Temperature ............... -20°C to +85°C
Operating Humidity .................. 0% to 90% non-condensing
Short-Circuit Protection ........... Shutdown; Auto-recovery
Overload Protection ................. Set at 110% to 160% (except
110% to 170% for 48V and up)
of the max. rating; Auto-recovery
Overvoltage Protection ............ Shutdown and latch off;
AC recycle to restart
Withstand Voltage ....................5,656 Vdc from input to output,
.............................................2,121 Vdc from input to ground
MTBF ........................................ 155K hours at full load and
25°C ambient temperature
calculated per MIL-HDBK-217F
*The output regulation will go up to ±10% at -20°C to -0°C ambient temperature.
MODELS LIST
Product No. 1
Output Voltage Output Current
Regulation
Max. Output Power Ripple & Noise2
TWM40-05
5V
0-8A
±3%
40W
100mV
TWM65-12
12V
0-5.416A
±3%
65W
100mV
TWM65-15
15V
0-4.333A
±3%
65W
100mV
TWM65-19
19V
0-3.421A
±3%
65W
100mV
TWM65-24
24V
0-2.708A
±3%
65W
120mV
TWM65-28
28V
0-2.321A
±3%
65W
150mV
TWM65-30
30V
0-2.167A
±3%
65W
150mV
TWM65-36
36V
0-1.806A
±3%
65W
160mV
TWM65-48
48V
0-1.355A
±3%
65W
160mV
TWM65-56
56V
0-1.161A
±3%
65W
200mV
Notes:
1. Use TWM65-xx.x to order a special output voltage of xx.xV between 12.1V and 56.9V.
2. R&N are peak-to-peak values measured at 20 MHz bandwidth with an 100µF electrolytic capacitor and a 0.1µF ceramic capacitor in parallel.
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Building 13 • Santa Clara, California 95054, USA • 408-988-6616 • sales@trumpower.com • www.TRUMPower.com