FEATURES
• 90-264 VAC Universal Input
• 0.9 to 0.95 Power Factor
• 200W Output @ Convection Cooling
• No Load Power Consumption < 500mW
• Meet Efficiency Level V Requirement
• Approved to IEC/EN/UL 60601-1 3rd Edition
ELECTRICAL SPECIFICATIONS
Input Voltage ........................ 90-264 VAC
Input Frequency ................... 47-63 Hz
Input Current ........................ 3A (rms) max. @ 115 VAC,
1.5A (rms) max. @ 230 VAC
Inrush Current ..................... 35A max. @ 115 VAC, 70A max.
@ 230 VAC at cold start +25°C
Earth Leakage Current ........ 228 µA max. @ 264 VAC, 63 Hz
Power Factor ....................... >0.95 @115 VAC, >0.9 @230 VAC
Output Power Ratings ......... See table
Output Voltage Regulation .. ±5%
Ripple & Noise .................... See table
STANDARDS & COMPLIANCES
EMC Standards ................... EN 60601-1-2, EN55011,
............................................. FCC Part 18 Class B
Safety Standards ................. IEC/EN/UL 60601-1 3rd Edition,
............................................. ANSI/AAMI ES 60601-1: 2005,
............................................. CAN/CSA C22.2 No. 60601-1:08
Agency Approvals ................ CSA, CUS, TUV, CE, CB, FCC
Other Compliances ............. RoHS, CEC
GENERAL SPECIFICATIONS
Efficiency .................................. >87% at full load, 115 VAC
Switching Frequency ............... 53-58 KHz
Hold-up Time ........................... >16 ms at full load, 115 VAC
Operating Temperature ........... 0°C to +60°C
Derating ................................... Derate linearly 2.5%/°C from
+41°C to +60°C
Temperature Coefficient .......... ±0.04%/°C
Operating Humidity .................. 10% to 95% non-condensing
Short-Circuit Protection ........... Auto-recovery
Overload Protection ................. Set at 110% to 160% of the
maximum rating, Auto-recovery
Overvoltage Protection ............ Shutdown and latch off, AC
............................................. recycle
Withstand Voltage .................... 4,000 VAC from input to output,
1,500 VAC from input to ground,
500 VAC from output to ground
MTBF ........................................ >140K hours at full load and
25°C ambient temperature
calculated per MIL-HDBK-217F
TWM200 Series
200W Medical Grade AC/DC Power Supplies
MODELS LIST
Product No.
Output
Voltage
Max. Load
Output Ripple & Noise
Max.
@ Convection Regulation
(Vp-p) (1)
Output Power
TWM200-12
12V
16.66A
±5%
240mV
200W
TWM200-16
16V
12.50A
±5%
300mV
200W
TWM200-19
19V
10.53A
±5%
300mV
200W
TWM200-20
20V
10A
±5%
300mV
200W
TWM200-24
24V
8.33A
±5%
300mV
200W
TWM200-36
36V
5.555A
±5%
300mV
200W
TWM200-48
48V
4.166A
±5%
300mV
200W
Notes:
1. R&N are measured at 20 MHz bandwidth with a 10µF electrolytic capacitor and
a 0.1µF ceramic capacitor in parallel.
MECHANICAL SPECIFICATIONS
Top View
176.0
6.0
7.5
95.0
+
-
23.0
50.0 47.0
20.0
6.0
7.5
176.0
164.0
95.0 80.0
6X M3 PEM Nuts on the bottom and side of
U-bracket for mounting
20.0
6.0
7.5
Notes:
1. Dimensions in mm
2. Tolerance: ±0.5mm
Bottom View