English
Language : 

TSM15 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – Medical Grade 15W Desktop Power Adapters
TRUMPower
TSM15 Series
Medical Grade 15W Desktop Power Adapters
FEATURES
• IEC 320/C14 Inlet
• Class I with 2 MOPP
• LED Power Indicator
• 5V to 36V Output Voltages
• 3,000 Meters Operating Altitude
• 0.3W No Load Power Consumption
• UL, cUL, TUV, PSE, FCC, CB & CE Approvals
INPUT SPECIFICATIONS
Input Voltage Range ................. 90~275 VAC
Input Frequency ........................ 47~63Hz
Input Current ............................. 0.32A rms max. @ 100 VAC,
0.19A rms max. @ 240 VAC
Inrush Current .......................... 23A @ 100 VAC or 55A
@ 240 VAC, at 25°C cold start
No Load Power Consumption... 0.3W max.
Earth Leakage Current ............. 100µA max. @ 240 VAC, 60 Hz
OUTPUT SPECIFICATIONS
Output Power Ratings .............. See models list
Minimum Load ......................... Not required
Line Regulation ........................ ±1% max. @ full load
Total Regulation ....................... See models list
Ripple & Noise ......................... See models list
Overload Protection ................. Set at 110-150% of its nominal
output voltage; Auto recovery
Overvoltage Protection .............. Set at 112-132% of its nominal
output voltage
Temperature Coefficient .......... ±0.04%/°C max.
Transient Response ................ 4ms @ 110 VAC, full load
GENERAL SPECIFICATIONS
Efficiency .................................. See models list
Hold-up Time ............................ 10ms min. @ 100 VAC, full load
Start-up Time ............................ 1 second max. @ full load
Operating Temperature ........... -10°C to +70°C
Derating .................................... Derate from 100% at +50ºC
linearly to 50% at +70°C
Storage Temperature ............... -40°C to +85°C
Relative Humidity ..................... 0% to 95% non-condensing
Withstand Voltage .................... 4,000 VAC from input to output,
.................................................. 1,500 VAC from input to ground
Operating Altitude ..................... 3,000 meters max.
MTBF ......................................... 100K hours minimum at full load,
25°C ambient, calculated per
MIL-HDBK-217F
STANDARDS & COMPLIANCES
IEC/EN 60601-1-2 ............... EMC & Immunity Performance
EN 55011, CISPR11 ........... Class B, conducted & radiated
FCC, VCCI ........................... Class B, conducted & radiated
EN 61000-3-2 ...................... Harmonic distortion, Class B & D
EN 61000-3-3 ...................... Line flicker
EN 61000-4-2 ...................... ESD, ±8 KV air and ±6 KV contact
EN 61000-4-3 ...................... Radiated immunity, 3V/m
EN 61000-4-4 ...................... Fast transient/burst, ±2 KV
EN 61000-4-5 ...................... Surge, ±1 KV diff., ±2 KV com.
EN 61000-4-6 ...................... Conducted immunity, 3 Vrms
EN 61000-4-8 ...................... Magnetic field immunity, 3A/m
EN 61000-4-11 .................... Voltage dips,
30% reduction for 500ms,
60% reduction for 100ms,
>95% reduction for 10ms
Safety Standards ................. IEC/EN/UL 60601-1 (3rd Ed.),
............................................. ANSI/AAMI ES 60601-1 (1st Ed.),
............................................. CSA C22.2 No. 60601-1 (2nd Ed.)
Agency Approvals ................ UL, cUL, TUV, PSE, FCC, CE, CB
Other Compliances ............ RoHS2, CEC V
MECHANICAL SPECIFICATIONS
Unit: inch. [mm]
Tolerance: 0.02 [0.5] max.
Weight: 170g (0.38 lb) approx.
Notes:
1. Standard output connector: 2.5 x 5.5 mm or 2.1 x 5.5 mm barrel plug
2. Standard output cable: 4ft, AWG#16 for 5-7.99V output models;
4ft, AWG#18 for 8-36V output models.
3. Contact TRUMPower for other output cable and connector options.
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Building 13 • Santa Clara, California 95054, USA • 408-988-6616 • sales@trumpower.com • www.TRUMPower.com