English
Language : 

TMW10_16 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – 6.6 to 15 Watts Medical Grade Power Adapters
TRUMPower
FEATURES
• Single 3.3V to 24V DC Outputs
• Interchangeable and Fixed AC Plugs
• USB Socket on 5V Output Models
• Approved to IEC/EN/UL 60601-1 Edition 3.1
• Meets IEC/EN 60601-1-2 4th Edition Medical EMC
• DOE/GEMS Level VI & ErP Stage II Compliant
• Black (Standard) and White (Optional) Colors
TMW10/12/15 Series
6.6 to 15 Watts Medical Grade Power Adapters
INPUT SPECIFICATIONS
Input Voltage Range ............... 90-264 VAC
Input Frequency ...................... 47-63 Hz
Input Current ........................... 0.32A rms @ 115 VAC,
................................................ 0.19A rms @ 230 VAC
Inrush Current ........................ 60A @ 240 VAC, at 25°C cold start
Touch Current ......................... 100 µA max.
No Load Input Power. ............. 0.1W max.
................................................
OUTPUT SPECIFICATIONS
Output Power Ratings ............ See table
Line Regulation ...................... ±1% max.
Load Regulation ..................... ±5% max.
Ripple and Noise* .................. 1% peak to peak max.
Over Voltage Protection .......... Shut down
Short Circuit Protection .......... Shut down; Auto-recovery
Over Current Protection ......... Shutdown, Auto-recovery
Temperature Coefficient ........ ±0.04%/°C max.
Transient Response .............. 0.5 ms @ 50% load change
* Measured with 20MHz bandwidth at rated line voltage and output load ranges, with
a 10µF tantalum capacitor in parallel with a 0.1µF ceramic capacitor in parallel
across the output.
STANDARDS & COMPLIANCES
IEC/EN 60601-1-2:2014 ....... EMC & Immunity Performance
EN 55011, CISPR 11, FCC ... Class B, conducted & radiated
EN 61000-3-2 ........................ Harmonic distortion, Class A
EN 61000-3-3 ........................ Line fluctuations & flicker
EN 61000-4-2 ........................ ESD, ±15 KV air and ±8 KV contact
EN 61000-4-3 ........................ Radiated immunity, 10V/m
EN 61000-4-4 ........................ Fast transient/burst, ±2 KV
EN 61000-4-5 ........................ Surge, ±1 KV diff., ±2 KV com.
EN 61000-4-6 ........................ Conducted immunity, 6 Vrms
EN 61000-4-8 ........................ Magnetic field immunity, 30A/m
EN 61000-4-11 ...................... Voltage dip immunity,
30% reduction for 500ms,
60% reduction for 100ms,
>95% reduction for 10ms
Safety Standards ................... UL/IEC/EN 60601-1 (Edition 3.1),
............................................... ANSI/AAMI ES 60601-1 (2012),
............................................... CSA C22.2 No. 60601-1 (2014);
IEC/EN 60950-1 (2nd Edition, CB)
Pending Agency Approvals .... UL, cUL, CE, CB, TUV/GS
Other Compliances .............. RoHS, DOE/GEMS Level VI,
............................................... ErP Stage II
GENERAL SPECIFICATIONS
Efficiency ................................ 78% min. @ full load
Hold-up Time .......................... 10 ms @ full load
Operating Temperature ......... 0°C to +40°C
Storage Temperature ............. -20°C to +80°C
Operating Humidity ................ 20% to 80% non-condensing
Storage Humidity .................... 10% to 90% non-condensing
Withstand Voltage .................. 4,000 VAC, input-output
MTBF ....................................... 300 KHrs minimum at full load,
25°C ambient, calculated per
Telecordia SR-332
MODELS LIST
Product No.
Output Maximum
Voltage Current
Maximum
Power
TMW10-3.3-X
3.3V
2A
6.6W
TMW10-5-X
5V
2A
10W
TMW15-5-X
5V
3A
15W
TMW10-9-X
9V
1.2A
10.8W
TMW12-12-X
12V
1A
12W
TMW12-15-X
15V
0.8A
12W
TMW12-18-X
18V
0.67A
12W
TMW12-24-X
24V
0.5A
12W
Note: X = AC plug code. See the tables and drawings below for details.
Code X Designation/Description
U
Fixed US Plug
E
Fixed EU Plug
B
Fixed UK Plug
I
Interchangeable "Main Body"
without the AC Plug
Interchangeable Plug Part No.
US Plug
TIPU
EU Plug
TIPE
UK Plug
TIPB
AU Plug
TIPA
Notes:
1. Specify the AC plug code when ordering, e.g. TMW10-5-U for 5V output with US
AC plug, TMW12-12-I for 12V output "main body" for interchangeable plug, etc.
2. For interchangeable type, at least one interchangeable plug must be ordered
together with the "main body" (X=I), e.g. TMW10-5-I for 5V output main body and
TIPU for US type interchangeable plug, etc.
3. Add suffix "P" to order units with a built-in USB socket, e.g. TMW10-5-UP for 5V
output with a fixed US plug and USB type A receptacle. The maximum current for
USB type A connector is 2.4A.
4. Standard color is black. Add suffix "W" for white version, e.g. TMW10-5-UPW.
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Building 13 • Santa Clara, California 95054, USA • 408-988-6616 • sales@trumpower.com • www.TRUMPower.com