English
Language : 

TMP55 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – 55 WATT MEDICAL POWER ADAPTERS
TRUMPower
55 WATT MEDICAL POWER ADAPTERS
DESCRIPTION
The TMP55 series consists of 15 standard
desktop AC/DC switching power adapters with
40 to 55 watts of continuous output power. The
units are enclosed in a 94V-1 rated plastic case
with an IEC 320 inlet to mate with
interchangeable cord for world-wide use. An
on/off switch can be added at customer’s
request. All models meet CISPR 11 and FCC
class B emission limits, and are designed for
medical applications.
TMP55 SERIES
(MDD)
FEATURES
‧ 15 standard desktop models
‧ Single, dual or triple outputs
‧ Optional output connectors
‧ Optional on/off switch
‧ 100% burn-in
‧ Wide input range 85 to 264VAC
‧ Input surge current protection
‧ Overvoltage protection
‧ Overcurrent protection
INPUT SPECIFICATIONS
Input voltage: 85 to 264 VAC
Input frequency: 47 to 63 Hz
Input current: 1.50A (rms) for 115 VAC
0.90A (rms) for 230VAC
Leakage current: 90µA max. @ 115VAC, 60Hz
150µA max. @ 230VAC, 50Hz
Safety Standard Approvals:
UL 2601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
OUTPUT SPECIFICATIONS
Output voltage/current: See rating chart
Total output power:
55 watts maximum
Ripple and noise:
1% peak to peak maximum
Overvoltage protection: Provided on output #1 only;
set at 112-132% of its
nominal output voltage
Overcurrent protection: All outputs protected to short
circuit conditions
Temperature coefficient: All outputs ±0.04%/ºC
maximum
Transient response:
Maximum excursion of 4% or
better on all models,
recovering to 1% of final
value within 500µs after a
25% step load change
ENVIRONMENTAL SPECIFICATIONS
O p e r a t i n g t e m p e r a t u r e:0 º C t o + 7 0 º C
Storage temperature: -40ºC to +85ºC
Relative humidity:
Derating:
5% to 95%
non-condensing
Derate from 100% at +50ºC
linearly to 50% at +70ºC
GENERAL SPECIFICATIONS
Switching frequency: 40KHz ±5KHz
Efficiency:
Hold-up time:
Line regulation:
70% minimum on single
output models with Vo ≥10V,
65% minimum on the others
10 msec minimum at 110VAC
±0.5% maximum at full load
Inrush current:
15 amps @ 115VAC, or 30 amps
@230VAC, at 25ºC cold start
Withstand voltage: 4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
MTBF:
450,000 hours minimum at full load at
25ºC ambient, calculated per
MIL-HDBK-217F
EMC performance (EN60601-1-2: 2001)
EN55011:
FCC:
VCCI:
Class B conducted, Class B radiated
Class B conducted, Class B radiated
Class B conducted, Class B radiated
EN61000-3-2:
EN61000-3-3:
EN61000-4-2:
Harmonic distortion, Class A
Line flicker
ESD, ±8KV air and ±6KV contact
EN61000-4-3:
Radiated immunity, 3V/m for 80~2500MHz
EN61000-4-4:
Fast transient/burst, ±2KV
EN61000-4-5:
Surge, ±1KV diff, ±2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
EN61000-4-8:
Magnetic field immunity, 3A/m
E N 6 1 0 0 0 - 4 - 11 : Voltage dips, 30% reduction for 500ms,
60% reduction for 100ms and >95%
reduction for 10ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: sales@trumpower.com ● Website: www.TRUMPower.com