DESCRIPTION
The TMP30 series of AC/DC switching power
adapters are capable of delivering 25 to 30 watt
output power. They are enclosed in a 94V-1
rated polyphenylene-oxide case with an IEC320
inlet to mate with interchangeable cord for
world-wide use. All models meet CISPR 11 and
FCC class B emission limits, and are designed
for medical applications.
TMP30 SERIES
(MDD)
F E AT U R E S
‧ 10 standard desktop models.
‧ Optional output connectors
‧ Optional on/off switch
‧ 100% burn-in at full rated load
‧ 90VAC to 264 VAC universal input
‧ Input surge current protection
‧ Overvoltage protection
‧ Overcurrent protection
‧ Short-circuit protection with auto-recovery
INPUT SPECIFICATIONS
Input voltage:
Input frequency:
Input current:
Leakage current:
85 to 264VAC
47 to 63Hz
0.80A( rms) for 115VAC
0.50A( rms) for 230VAC
90µA max. @ 115VAC, 60Hz
150µA max. @ 230VAC, 50Hz
OUTPUT SPECIFICATIONS
Output voltage/current:
Output power range:
Ripple and noise:
Overvoltage protection:
See rating chart
30 watts maximum
1% peak to peak maximum
Set at 112-132% of the higher
extreme of its nominal output
voltage
Overcurrent protection: The output is protected to
short circuit conditions
Temperature coefficient: All outputs ±0.04%/ºC
maximum
Transient response:
Maximum excursion of 4% or
better on all models,
recovering to 1% of final
value within 500µs after a
25% step load change
ENVIRONMENTAL SPECIFICATIONS
O p e r a t i n g t e m p e r a t u r e:0 º C t o + 7 0 º C
Storage temperature: -40ºC to +85ºC
Relative humidity:
5% to 95% non-condensing
Derating:
Derate from 100% at +50ºC
linearly to 50% at +70ºC
Safety Standard Approvals:
UL 2601-1,
CSA C22.2 No. 601.1
TÜV EN60601-1
GENERAL SPECIFICATIONS
Switching frequency: 70KHz ±5KHz
Efficiency:
Hold-up time:
70% minimum on all models with
output voltage ≥7V or 65% minimum
on the others
20ms minimum at 115VAC
Line regulation:
±0.5% maximum at full load
Inrush current:
15A @ 115VAC, or 30A @ 230VAC, at
25ºC and cold start
Withstand voltage: 4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
MTBF:
400,000 hours minimum at full load at
25ºC ambient, calculated per
MIL-HDBK-217F
EMC Performance (EN60601-1-2: 2001)
EN55011:
Class B conducted, class B radiated
FCC:
Class B conducted, class B radiated
VCCI:
Class B conducted, class B radiated
EN61000-3-2: Harmonic distortion, class A and D
EN61000-3-3: Line flicker
EN61000-4-2:
EN61000-4-3:
EN61000-4-4:
EN61000-4-5:
EN61000-4-6:
EN61000-4-8:
EN61000-4-11:
ESD, ±8KV air and ±6KV contact
Radiated immunity, 3V/m for 80~2500MHz
Fast transient/burst, ±2KV
Surge, ±1KV diff, ±2KV com.
Conducted immunity, 3Vrms
Magnetic field immunity, 3A/m
Voltage dips, 30% reduction for 500ms,
60% reduction for 100ms and >95%
reduction for 10ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: sales@trumpower.com ● Website: www.TRUMPower.com