English
Language : 

TMP180 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – 180 WATT MEDICAL POWER ADAPTERS
TRUMPower
180 WATT MEDICAL POWER ADAPTERS
TMP180 SERIES
Class I, Class II
DESCRIPTION
The TMP180 series of AC/DC switching power supplies
delivers 180 watts of continuous output power. They are
enclosed in a 94V-1 rated polyphenylene-oxide case with
AC inlet of the IEC 320/C14 or IEC320/C8 to mate with an
interchangeable cord for world-wide use. All models meet
CISPR 11 and FCC class B emission limits, and are designed
for medical applications, not for life- supporting equipment.
FEATURES
High efficiency
Low ripple & noise
Overvoltage protection
Short-circuit protection
Overpower protection
Over temperature protection
100% burn-in at full rated load
Standby consumption less than 0.5 W
Compliant with CEC and ENERGY STAR efficiency
level V requirements
Compliant with RoHS requirements
0.95 Power factor
Safety Standard Approvals
UL 60601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
Note: Agency approvals ready on 24V output; pending on other outputs
INPUT SPECIFICATIONS
Input voltage:
Input frequency:
Input current:
Earth leakage current:
90-264 VAC
47-63 Hz
2.3 A (rms) for 100 VAC
0.8 A (rms) for 240 VAC
300 uA max. @ 264 VAC, 60 Hz
OUTPUT SPECIFICATIONS
Output voltage/current:
Total output power:
Ripple and noise:
Overvoltage protection:
Overcurrent protection:
Transient response:
See rating chart
See rating chart
380 mV peak to peak maximum
Set at 130% to 150% of its nominal
output voltage
All models protected to short circuit
conditions
+10% maximum excursion
Output load step is:
(1) 0% to 50%
(2) 50% to 100%
Slew rate = 0.5 A/uS
Frequency is 100 Hz and 1 KHz
ENVIRONMENTAL SPECIFICATIONS
Operating temperature:0ºC to +40ºC
Storage temperature: -20ºC to +80ºC
Relative humidity: 10% to 90% non-condensing
GENERAL SPECIFICATIONS
Hold-up time:
5 msec minimum at 100 VAC
Turn on delay time: 3 S maximum at 100 VAC
Power Factor:
0.95 typical
Efficiency:
87% min. at 100 VAC or 240 VAC
Line regulation:
±0.5% maximum at full load
Inrush current:
No damage
Withstand voltage: 2000 VAC from input to output
MTBF:
100,000 hours at full load at 25ºC ambient,
calculated per MIL-HDBK-217F
EMC Performance (IEC 60601-1-2)
EN55011:
FCC:
VCCI:
EN61000-3-2:
EN61000-3-3:
EN61000-4-2:
EN61000-4-3:
EN61000-4-4:
EN61000-4-5:
EN61000-4-6:
EN61000-4-8:
EN61000-4-11:
Class B conducted, Class B radiated
Class B conducted, Class B radiated
Class B conducted, Class B radiated
Harmonic distortion, Class D
Line flicker
ESD, ±15 KV air and ±8 KV contact
Radiated immunity, 3 V/m
Fast transient/burst, ±2 KV
Surge, ±1 KV diff., ±2 KV com.
Conducted immunity, 3 Vrms
Magnetic field immunity, 3 A/m
Voltage dips, 30% reduction for 500 ms, 60%
reduction for 100 ms and >95% reduction for 10
ms
Phone: 408-988-6616
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Fax: 408-988-6622 email: sales@trumpower.com Website: www.TRUMPower.com