TMP150 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – 150 WATT MEDICAL POWER ADAPTERS | |||
|
TRUMPower
150 WATT MEDICAL POWER ADAPTERS
DESCRIPTION
The TMP150 is a series of medical grade, class I
and class II, AC/DC desktop power adapters of 132
to 150 watts output power. They are enclosed in a
94V-1 rated case with an IEC 320/C14 or IEC
320/C18 AC inlet to mate with interchangeable cord
for world-wide use. All models meet CISPR 11 and
FCC class B emission limits, and are designed for
medical applications.
TMP150 SERIES
Class I, Class II
( LV D )
RoHS
FEATURES
⧠Low safety ground leakage current
⧠Optional output connectors
⧠100% burn-in
⧠90 to 264VAC wide input range
⧠Overvoltage & overcurrent protections
⧠0.98 PFC
⧠Compliant with CEC and Energy Star efficiency
level IV requirements
⧠No load consumption less than 0.5W
⧠Average active efficiency⧠85%
⧠Compliant with RoHS requirements
Safety Standard Approvalsï¼
UL 60601-1
CSA C22.2 No. 601.1
TÃV EN60601-1
INPUT SPECIFICATIONS
Input voltageï¼
90 to 264VAC
Input frequencyï¼
47 to 63Hz
Input currentï¼
2.0A (rms) for 115VAC
Earth leakage currentï¼
1.0A (rms) for 230VAC
90µA max. @ 115VAC, 60Hz
(Class I)
150µA max. @ 230VAC, 50Hz
E n c l o s u r e l e a k a g e c u r r e n t : 90µA max. @ 264VAC, 50Hz
(Class II)
OUTPUT SPECIFICATIONS
Output voltage/currentï¼
See rating chart
Total output powerï¼
See rating chart
Ripple and noiseï¼
2% peak to peak maximum
Overvoltage protectionï¼
Set at 112 â 140% of its nominal
output voltage
Overcurrent protectionï¼
All outputs protected to short
circuit conditions
Temperature coefficientï¼ All outputs ±0.04%/ºC maximum
Transient responseï¼
Maximum excursion of 4% or
better on all models, recovering
to 1% of final value within 500µs
after a 25% step load change
ENVIRONMENTAL SPECIFICATIONS
O p e r a t i n g t e m p e r a t u r e ï¼ 0 ºC t o + 6 0 ºC
Storage temperatureï¼
- 4 0 ºC t o + 8 5 º C
Relative humidityï¼
5% to 95% non-condensing
Deratingï¼
Derate from 100% at +40ºC linearly
to 50% at +60ºC
GENERAL SPECIFICATIONS
Switching frequencyï¼ 30KHz ~ 110KHz
Power factorï¼
Efficiencyï¼
Hold-up timeï¼
Line regulationï¼
0.98 typical at 115VAC
85% min.
15 msec minimum at 110VAC
±0.5% maximum at full load
Inrush currentï¼
Withstand voltageï¼
60 amps @ 115VAC or 100 amps
@ 230VAC, at 25ºC cold start
4000VAC from input to output
MTBFï¼
1500VAC from input to ground
500VAC from output to ground
150,000 hours minimum at full load
EMC Performance
At 25ºC ambient, calculated per
MIL-HDBK-217F
EN55011:
Class B conducted, Class B radiated
FCC:
VCCI:
EN61000-3-2:
Class B conducted, Class B radiated
Class B conducted, Class B radiated
Harmonic distortion, Class D
EN61000-3-3:
EN61000-4-2:
Line flicker
ESD, ±8KV air and ±6KV contact
EN61000-4-3:
Radiated immunity, 3V/m
EN61000-4-4:
Fast transient/burst, ±2KV
EN61000-4-5:
Surge, ±1KV diff, ±2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
EN61000-4-8:
Magnetic field immunity, 3A/m
EN61000-4-11:
Voltage dips, 30% reduction for
500ms, 60% reduction for 100ms
and >95% reduction for 10ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 â Fax: 408-988-6622 â email: sales@trumpower.com â Website: www.TRUMPower.com
|
▷ |