English
Language : 

TMP130 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – 130 WATT MEDICAL POWER ADAPTERS
TRUMPower
130 WATT MEDICAL POWER ADAPTERS
DESCRIPTION
The TMP130 series of AC/DC switching power
adapters are for 100 to 130 watts of continuous
output power. They are designed for medical
applications that require external power. The units
are enclosed in a 94V-1 rated polyphenylene-oxide
case with an IEC 320 inlet to mate with
interchangeable cord for world-wide use. An on/off
switch can be added at request. All models meet
CISPR 11 and FCC class B emission limits, and
comply with UL, CSA, IEC and CE requirements.
TMP130 SERIES
(MDD+EMC)
F E AT U R E S
‧ In compliance with EN61000-3-2, EN61000-3-3 and
EN61000-1-1-2
‧ Optional output connectors
‧ Optional on/off switch
‧ Optional Power Fail Detect (PFD) signal
‧ 100% burn-in at full load
‧ Auto-ranging input
‧ Input surge current protection
‧ Overvoltage protection
‧ Thermal protection
Safety Standard Approvals:
UL 2601-1
CSA C22.2 No. 601.1
INPUT SPECIFICATIONS
Input voltage: 90 to 132VAC or 180 to 264VAC
(auto-ranging)
Input frequency: 47 to 63 Hz
Input current: 3.0A (rms) for 115 VAC
1.7A (rms) for 230VAC
Leakage current: 90µA max. @ 115VAC, 60Hz
150µA max. @ 230VAC, 50Hz
OUTPUT SPECIFICATIONS
Output voltage/current: See rating chart
Total output power:
130 watts maximum
Ripple and noise:
1% peak to peak maximum
Overvoltage protection: Provided on output: set at
112 – 132% of its nominal
output voltage
Overcurrent protection: All outputs protected to
short circuit conditions
Temperature coefficient: All outputs ±0.04%/ºC
maximum
Transient response:
Maximum excursion of 4% or
better on all models,
recovering to 1% of final
value within 500µs after a
25% step load change
ENVIRONMENTAL SPECIFICATIONS
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
0ºC to +70ºC
-40ºC to +85ºC
5% to 95% non-condensing
Derate from 100% at +50ºC
linearly to 50% at +70ºC
TÜV EN60601-1
GENERAL SPECIFICATIONS
Switching frequency: 68 Khz ±8Khz
Efficiency:
70% minimum on Vo ≥12V
65% minimum on others
Hold-up time:
10 msec minimum at 115VAC
Line regulation:
±0.5% maximum at full load
Inrush current:
15amps @115VAC, or 30amps
@230VAC, at 25ºC and cold start
Withstand voltage: 4000VAC from input to output
1500VAC from input to ground
5000VAC from output to ground
MTBF:
400,000 hours minimum at full load
at 25ºC ambient, calculated per
MIL-HDBK-217F, excluding DC fan.
EMC Performance (EN60601-1-2: 2001)
EN55011:
Class B conducted, class B radiated
FCC:
Class B conducted, class B radiated
VCCI:
Class B conducted, class B radiated
EN61000-3-2:
Harmonic distortion, class A
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD, ±8KV air and ±6KV contact
EN61000-4-3:
Radiated immunity, 3V/m for
80~2500MHz
EN61000-4-4:
Fast transient/burst, ±2KV
EN61000-4-5:
Surge, ±1KV diff, ±2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
EN61000-4-8:
Magnetic field immunity, 3A/m
EN61000-4-11:
Voltage dips, 30% reduction for
500ms, 60% reduction for 100ms and
>95% reduction for 10ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: sales@trumpower.com ● Website: www.TRUMPower.com