English
Language : 

TMP120 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – 120 WATT MEDICAL POWER ADAPTERS
TRUMPower
120 WATT MEDICAL POWER ADAPTERS
DESCRIPTION
The TMP120 series of AC/DC switching power
adapters are for 96 to 120 watts of continuous
output power. They are enclosed in a 94V-1 rated
case with an IEC 320/C14 or C18 inlet to mate with
interchangeable AC power cord for world-wide
use. All models meet CISPR 11 and FCC class B
emission limits, and are designed for medical
applications.
TMP120 SERIES
Class I, Class II
(LVD)
RoHS
FEATURES
‧ Low safety ground leakage current
‧ Optional output connectors
‧ 100% burn-in
‧ 90 to 264VAC wide input range
‧ Overvoltage and overcurrent protections
‧ Optional IEC 320/C18 inlet with class II insulation
‧ Compliant with CEC and Energy Star efficiency
level IV requirements
‧ No load power consumption less than 0.5W
‧ Average active efficiency ≥85%
‧ Compliant with RoHS requirements
Safety Standard Approvals:
UL 60601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
INPUT SPECIFICATIONS
Input voltage:
90 to 264VAC
Input frequency:
47 to 63Hz
Input current:
1.60A (rms) for 115 VAC
0.80A (rms) for 230 VAC
Earth leakage current:
(Class I)
90µA max. @ 115 VAC/60Hz
150µA max. @ 230 VAC/50Hz
E n c l o s u r e l e a k a g e c u r r e n t : 90µA max. @ 264 VAC/50Hz
(Class II)
OUTPUT SPECIFICATIONS
Output voltage/current: See rating chart
Total output power:
120 watts maximum
Ripple and noise:
2% peak to peak maximum
Overvoltage protection: Provided on output #1 only;
set at 112 -140% of its nominal
output voltage
Overcurrent protection: All outputs protected to short
circuit conditions
Temperature coefficient: All outputs ±0.04%/ºC maximum
Transient response:
Maximum excursion of 4% or
better on all models, recovering
to 1% of final value within 500µs
after a 25% step load change
GENERAL SPECIFICATIONS
Switching frequency: 50KHz ~ 110KHz
Power factor:
0.98 typical at 115VAC
Efficiency:
85% min.
Hold-up time:
15 msec minimum at 115VAC
Line regulation:
±0.5% maximum at full load
Inrush current:
Withstand voltage:
MTBF:
60 amps @ 115VAC or 90 amps
@230VAC, at 25ºC cold start
4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
150,000 hours minimum at full load at
25ºC ambient, calculated per
MIL-HDBK-217F
EMC Performance (EN60601-1-2: 2001)
EN55011:
FCC:
Class B conducted, Class B radiated
Class B conducted, Class B radiated
VCCI:
EN61000-3-2:
EN61000-3-3:
Class B conducted, Class B radiated
Harmonic distortion, Class A and D
Line flicker
EN61000-4-2:
ESD, ±8KV air and ±6KV contact
EN61000-4-3:
Radiated immunity, 3V/m
EN61000-4-4:
Fast transient/burst, ±2KV
EN61000-4-5:
Surge, ±1KV diff, ±2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
ENVIRONMENTAL SPECIFICATIONS
O p e r a t i n g t e m p e r a t u r e : 0 ºC t o + 6 0 ºC
S t o r a g e t e m p e r a t u r e : - 4 0 ºC t o + 8 5 ºC
Relative humidity:
5% to 95% non-condensing
Derating:
Derate from 100% at +40ºC
EN61000-4-8:
EN61000-4-11:
Magnetic field immunity, 3A/m
Voltage dips, 30% reduction for 500ms,
60% reduction for 100ms and >95%
reduction for 10ms
linearly to 50% at +60ºC
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: sales@trumpower.com ● Website: www.TRUMPower.com