|
TMP110 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – 110 WATT MEDICAL POWER ADAPTERS | |||
|
TRUMPower
DESCRIPTION
The TMP110 series of AC/DC switching power
adapters are for 110 watts of continuous
output power. They are enclosed in a 94V-1
rated polyphenylene-oxide case with an IEC
320 inlet to mate with interchangeable cord for
world-wide use. An on/off switch can be added
at request. All models meet CISPR 11 and FCC
class B emission limits, and are designed for
medical applications.
110 WATT MEDICAL POWER ADAPTERS
TMP110 SERIES
(MDD)
F E AT U R E S
⧠Dual, triple or quadruple outputs
⧠Optional output connectors
⧠Optional on/off switch
⧠Optional Power Fail Detect (PFD) signal
⧠100% burn-in
⧠Wide input range 85 to 264 VAC
⧠Input surge current protection
⧠Overvoltage protection
⧠Overcurrent protection
Safety Standard Approvalsï¼
UL 2601-1
CSA C22.2 No. 601.1
TÃV EN60601-1
INPUT SPECIFICATIONS
Input voltageï¼ 85 to 264 VAC
Input frequencyï¼ 47 to 63 Hz
Input currentï¼ 3.20A (rms) for 115 VAC
1.80A (rms) for 230VAC
Leakage currentï¼ 90µA max. @ 115VAC, 60Hz
150µA max. @ 230VAC, 50Hz
ENVIRONMENTAL SPECIFICATIONS
Operating temperatureï¼ 0ºC to +70ºC
Storage temperatureï¼
-40ºC to +85ºC
Relative humidityï¼
Deratingï¼
5% to 95% non-condensing
Derate from 100% at +50ºC
linearly to 50% at +70ºC
OUTPUT SPECIFICATIONS
Output voltage/currentï¼ See rating chart
Total output powerï¼
110 watts maximum
Ripple and noiseï¼
Overvoltage protectionï¼
1% peak to peak maximum
Provided on output #1 only;
set at 112-132% of its nominal
output voltage
Overcurrent protectionï¼ All outputs protected to short
circuit conditions
Temperature coefficientï¼ All outputs ±0.04%/ºC
maximum
Transient responseï¼
Maximum excursion of 4% or
better on all models,
recovering to 1% of final
value within 500us after a
25% step load change
PFD signal (optional):
TTL logic high for normal
operation and TTL logic low
upon loss of input power. This
signal appears at least 1ms
prior to +5V output dropping
5% below its nominal value.
This signal also provides a
minimum delay of 100ms after
+5V is within regulation.
GENERAL SPECIFICATIONS
Switching frequencyï¼ 20KHz to 250KHz, varies with load and line
Efficiencyï¼
Hold-up timeï¼
Line regulationï¼
70% minimum on single output models with
Vo â¥12V, 65% minimum on the others
12 msec minimum at 110VAC
±0.5% maximum at full load
Inrush currentï¼
15 amps @ 115VAC, or 30 amps
@230VAC, at 25ºC cold start
Withstand voltageï¼ 4000VAC from input to output 1500VAC
from input to ground
500VAC from output to ground
MTBFï¼
550,000 hours minimum at full load at 25ºC
ambient, calculated per MIL-HDBK-217F,
excluding DC fan.
EMC Performance (EN60601-1-2: 2001)
EN55011:
Class B conducted, Class B radiated
FCC:
Class B conducted, Class B radiated
VCCI:
Class B conducted, Class B radiated
EN61000-3-2:
Harmonic distortion, Class A
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD, ±8KV air and ±6KV contact
EN61000-4-3:
Radiated immunity, 3V/m for 80~2500MHz
EN61000-4-4:
Fast transient/burst, ±2KV
EN61000-4-5:
Surge, ±1KV diff, ±2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
EN61000-4-8:
Magnetic field immunity, 3A/m
EN61000-4-11:
Voltage dips, 30% reduction for 500ms, 60%
reduction for 100ms and >95% reduction for
10ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 â Fax: 408-988-6622 â email: sales@trumpower.com â Website: www.TRUMPower.com
|
▷ |