DESCRIPTION
The TMC60 series of compact, open PCB constructed, AC-DC
switching power supplies are capable of delivering 37.5-64
watts of continuous output power at convection cooling. They
operate at 90-264 VAC input voltage without the need of
voltage selection, and are suited for medical, information
technology and industrial applications. Approval to both EN
60601-1 and EN 60950-1 safety standards improves design-in
time and reduces end equipment compliance costs.
TMC60 SERIES
FEATURES
z Medical and ITE approvals
z Compact size 2” x 4” x 1.18”
z Single, dual and triple outputs
z Full range input 90-264 VAC
z Low earth leakage current
z Level B emissions
z RoHS compliant
INPUT SPECIFICATIONS
Input voltage:
90-264 VAC
Input frequency:
47-63 Hz
Input current:
1.3 A (rms) for 100 VAC
0.7 A (rms) for 240 VAC
Earth leakage current: 150 µA max. @ 264 VAC, 60 Hz
OUTPUT SPECIFICATIONS
Output voltage/current: See rating chart
Maximum output power: See rating chart
Ripple and noise:
2% peak to peak on 5.0 V model, 1% peak
to peak on other models.
Overvoltage protection: Provided on output #1 only;
set at 112-132% of its nominal output
voltage
Overcurrent protection: All outputs protected to short circuit
conditions
Temperature coefficient: All outputs ±0.04% /oC maximum
Transient response:
Maximum excursion of 4% or better on all
models, recovering to 1% of final value
within 500 µs after a 25% step load
change
ENVIRONMENTAL SPECIFICATIONS
Operating temperature:
Storage temperature:
-10oC to +70 oC
-40 oC to +85 oC
Relative humidity:
Derating:
5% to 95% non-condensing
Derate from 100% at +50 oC linearly to
50% at +70 oC
SAFETY STANDARD APPROVALS
UL 60601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
GENERAL SPECIFICATIONS
Switching frequency:
Efficiency:
Hold-up time:
Line regulation:
Inrush current:
Withstand voltage:
MTBF:
EMC Performance
EN55011 / EN55022:
FCC:
VCCI:
EN61000-3-2:
EN61000-3-3:
EN61000-4-2:
EN61000-4-3:
EN61000-4-4:
EN61000-4-5:
EN61000-4-6:
EN61000-4-8:
EN61000-4-11:
62±5 KHz
80-88% typical except TMC60-T31-3 and
TMC60-T31-5 at 75% typical
12 msec minimum at 110 VAC
±0.5% maximum at full load
30 A @ 115 VAC, or 60 A @ 230 VAC,
at 25 oC cold start
4000 VAC from input to output
1500 VAC from input to ground
500 VAC from output to ground
400,000 hours at full load at 25 oC ambient,
calculated per MIL-HDBK-217F
Class B conducted, class B radiated
Class B conducted, class B radiated
Class B conducted, class B radiated
Harmonic distortion, class A and D
Line flicker
ESD, ±8 KV air and ±6 KV contact
Radiated immunity, 3 V/m
Fast transient/burst, ±2 KV
Surge, ±1 KV diff., ±2 KV com.
Conducted immunity, 3 Vrms
Magnetic field immunity, 3 A/m
Voltage dips,
30% reduction for 500 ms
60% reduction for 100 ms
>95% reduction for 10 ms