English
Language : 

TMC40 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – 30-48 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES
TRUMPower
30-48 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES
DESCRIPTION
The TMC40 series of compact, open PCB constructed, AC-DC
switching power supplies are capable of delivering 30-48
watts of continuous output power at convection cooling. They
operate at 90-264 VAC input voltage without the need of
voltage selection, and are suited for medical, information
technology and industrial applications. Approval to both
EN60601-1 and EN60950-1 safety standards improves
design-in time and reduces end equipment compliance costs.
TMC40 SERIES
FEATURES
z Medical and ITE approvals
z Compact size 2” x4” x1.18”
z Single, dual and triple outputs
z Full-range input 90-264 VAC
z Low earth leakage current
z Level B emissions
z RoHS compliant
SAFETY STANDARD APPROVALS
UL 2601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
INPUT SPECIFICATIONS
Input voltage:
90-264 VAC
Input frequency:
47-63 Hz
Input current:
0.9 A (rms) for 100 VAC
0.5 A (rms) for 240 VAC
Earth Leakage current: 150 µA max. @ 264 VAC, 63 Hz
OUTPUT SPECIFICATIONS
Output voltage/current: See rating chart
Maximum output power: See rating chart
Ripple and noise:
2% peak to peak on 5.0 V model,
1% peak to peak on other models.
Overvoltage protection: Provided on output #1 only; set at
112–132% of its nominal output voltage
Overcurrent protection: All outputs protected to short circuit
conditions
Temperature coefficient: All outputs ±0.04% /°C maximum
Transient response:
Maximum excursion of 4% or better on all
models, recovering to 1% of final value
within 500 µs after a 25% step load
change
ENVIRONMENTAL SPECIFICATIONS
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
-10oC to +70oC
-40oC to +85oC
5% to 95% non-condensing
Derate from 100% to +50oC linearly to
50% at +70oC
GENERAL SPECIFICATIONS
Switching frequency:
Efficiency:
Hold-up time:
Line regulation:
Inrush current:
Withstand voltage:
MTBF:
EMC Performance
EN55011 / EN55022:
FCC:
VCCI:
EN61000-3-2:
EN61000-3-3:
EN61000-4-2:
EN61000-4-3:
EN61000-4-4:
EN61000-4-5:
EN61000-4-6:
EN61000-4-8:
EN61000-4-11:
62±5 KHz
80-88% typical except TMC40-T31-3 and
TMC40-T31-5 at 75% typical
12 msec minimum at 110 VAC
±0.5% maximum at fu6ll2lo±a5dKHz
25 A @ 115 VAC, or 5800A-8@8%23t0ypVicAaCl , except
at 25 oC cold start TMC40-T31-3
and
4000 VAC from input tToMoCut4p0u-tT31-5 at 75%
1500 VAC from input ttoypgicroaul nd
500 VAC from output to ground
400,000 hours at full load at 25 oC ambient,
calculated per MIL-HDBK-217F
Class B conducted, class B radiated
Class B conducted, class B radiated
Class B conducted, class B radiated
Harmonic distortion, class A and D
Line flicker
ESD, ±8 KV air and ±6 KV contact
Radiated immunity, 3 V/m
Fast transient/burst, ±2 KV
Surge, ±1 KV diff., ±2 KV com.
Conducted immunity, 3 Vrms
Magnetic field immunity, 3 A/m
Voltage dips,
30% reduction for 500 ms
60% reduction for 100 ms
>95% reduction for 10 ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: sales@trumpower.com ● Website: www.TRUMPower.com