English
Language : 

TMC150 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – 150 Watts AC/DC Medical & ITE Grade Power Supplies
TRUMPower
TMC150 Series
150 Watts AC/DC Medical & ITE Grade Power Supplies
FEATURES
• 2" x 4" x 1.2" Open Frame
• Low Leakage Current
• BF Rated Outputs
• Remote Sense & Optional PFD
• 0.5W No Load Power Consumption
MODELS LIST
GENERAL SPECIFICATIONS
Product No.(1)
Output
Voltage
Maximum Output Power (2)
@ Convection
@ 7.5 CFM
Current Wattage Current Wattage
TMC150-S12
12V
8.35A 100W 12.5A 150W
TMC150-S15
15V
6.7A 100W
10A
150W
TMC150-S18
18V
5.56A 100W 8.34A 150W
TMC150-S24
24V
4.2A
100W 6.25A 150W
TMC150-S30
30V
3.34A 100W
5A
150W
TMC150-S36
36V
2.78A 100W 4.17A 150W
TMC150-S48
48V
2.1A 100W 3.13A 150W
Notes:
1. Add suffix F to the P/N to order units with PFD, e.g. TMC150-S24F.
2. 168W peak power with 10% duty cycle maximum for less than 15 seconds with
the average output not to exceed the maximum power rating.
INPUT SPECIFICATIONS
Input Voltage Range .............. 90-264 VAC
Input Frequency ..................... 47-63 Hz
Input Current .......................... 1.7A rms @ 115 VAC,
0.85A rms @ 230 VAC
Inrush Current ....................... 80A @ 115 VAC or 160A @ 230 VAC,
at 25°C cold start
Earth Leakage Current .......... 105 µA max. @ 264 VAC, 63 Hz
OUTPUT SPECIFICATIONS
Output Power Ratings ........... See table
Minimum Load ...................... Not required
Tolerance ............................... ±2%
Ripple and Noise* ................. 1% Vp-p, 20 MHz bandwidth
Remote Sense ...................... Compensates cable losses up
to 0.5V
Overvoltage Protection .......... Set at 112-140% of its nominal
output voltage
Overcurrent Protection .......... Protected to short-circuit
conditions
Temperature Coefficient ....... ±0.04%/°C max.
Transient Response ............. Max. excursion of 4% or better on
all models, recovering to 1% of
final value within 500 µs after a
25% step load change
No Load Power Cons. .......... <0.5W w/o PFD, <1W with PFD
* Measured with a 10µF tantalum capacitor in parallel with a 0.1µF ceramic capacitor
at rated line voltage and load ranges.
Switching Frequency ............. 133 kHz typical
Power Factor ......................... 0.98 typical, with active PFC
Efficiency ............................... 87% typical @ 115 VAC,
89% typical @ 230 VAC
Hold-up Time ......................... 10 ms min. @ 120 VAC
Operating Altitude .................. 5,000 meters max.
Line Regulation ..................... ±0.5% max. @ full load
Operating Temperature ........ 0°C to +70°C
Derating ................................. Derate from 100% @ +50°C
linearly to 50% @ +70°C,
applicable to convection and
forced-air cooling conditions
Storage Temperature ............ -40°C to +85°C
Relative Humidity .................. 5% to 95% non-condensing
Withstand Voltage ................. 4,000 VAC, input-output (2 MOPP)
1,500 VAC, input-ground (1 MOPP)
1,500 VAC, output-ground
MTBF ...................................... 250 kHrs minimum at full load,
25°C ambient, calculated per
MIL-HDBK-217F
STANDARDS & COMPLIANCES
EN55011, EN55022 .......... Class B conducted, Class A radiated
FCC .................................... Class B conducted, Class A radiated
VCCI ................................... Class B conducted, Class A radiated
EN61000-3-2 ..................... Harmonic distortion, Class A & D
EN61000-3-3 ..................... Line flicker
EN61000-4-2 ..................... ESD, ±8 KV air and ±6 KV contact
EN61000-4-3 ..................... Radiated immunity, 3V/m
EN61000-4-4 ..................... Fast transient/burst, ±2 KV
EN61000-4-5 ..................... Surge, ±1 KV diff., ±2 KV com.
EN61000-4-6 ..................... Conducted immunity, 3 Vrms
EN61000-4-8 ..................... Magnetic field immunity, 3A/m
EN61000-4-11 ................... Voltage dips immunity,
30% reduction for 500ms,
60% reduction for 100ms,
>95% reduction for 10ms
Safety Standards ............... IEC/EN 60601-1 (3rd Edition),
ANSI/AAMI ES 60601-1:2005,
CSA C22.2 No. 60601-1:08;
UL/IEC/EN 60950-1 (2nd Edition),
CSA C22.2 No. 60950-1 (2nd Edition)
Agency Approvals .............. UL, cUL, TUV, CB, CE
Other Compliance ............. RoHS
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Building 13 • Santa Clara, California 95054, USA • 408-988-6616 • sales@trumpower.com • www.TRUMPower.com