English
Language : 

TMC100 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – 100 Watts Medical & ITE Switching Power Supplies
TRUMPower
TMC100 Series
100 Watts Medical & ITE Switching Power Supplies
FEATURES
• Compact Size 2" x 4" x 1.26"
• Ultra High Power Density 10W/cubic inch
• 90~132 VAC or 180~264 VAC Full Range Input
• Low Earth Leakage Current
• Conducted EMI Class B
INPUT SPECIFICATIONS
Input Voltage Range ............ 90~132 VAC or 180~264 VAC
Input Frequency ................... 47~63Hz
Input Current ........................ 1.9A rms @ 100~120 VAC
1.1A rms @ 200~240 VAC
Inrush Current ..................... 40A @ 115 VAC or 80A @ 230 VAC,
at 25°C cold start
Earth Leakage Current ........ 150µA max. @ 264 VAC, 63Hz
OUTPUT SPECIFICATIONS
Output Voltage/Current ........ See models list
Tolerance ............................. ±2%
Minimum Load .................... Not required
Ripple and Noise* ............... 150mV peak to peak on 5V model;
1% peak to peak on other models
Overvoltage Protection ........ Provided on output; set at 110-140%
of its nominal output voltage
Overcurrent Protection ........ All outputs protected to short-circuit
conditions
Temperature Coefficient ..... ±0.04%/°C max.
Transient Response ........... Max. excursion of 4% or better on
all models, recovering to 1% of
final value within 500µs after a
25% step load change
* Measured with 20MHz bandwidth at rated line voltage and output load ranges, with
a 10µF tantalum capacitor in parallel with a 0.1µF ceramic capacitor across the
output.
GENERAL SPECIFICATIONS
Switching Frequency ........... 100KHz
Efficiency ............................. 88~90% @ 230 VAC full load
Hold-up Time ....................... 12ms minimum at 110 VAC
Line Regulation ................... ±0.2% max. at full load
Operating Temperature ...... -10°C to +70°C
Derating ............................... Derate from 100% at +50°C linearly
to 50% at +70°C
Cooling ................................ Convection
Storage Temperature .......... -40°C to +85°C
Relative Humidity ................ 5% to 95% non-condensing
Withstand Voltage ............... 4000 VAC from input to output
1500 VAC from input to ground
500 VAC from output to ground
MTBF .................................... 270K hours minimum at full load,
25°C ambient, calculated per
MIL-HDBK-217F
MODELS LIST
Product No. Output Voltage Max. Current Max. Output Power
TMC100-S05
5V
20A
100W
TMC100-S12
12V
8.34A
100W
TMC100-S15
15V
6.7A
100W
TMC100-S18
18V
5.56A
100W
TMC100-S24
24V
4.2A
100W
TMC100-S28
28V
3.58A
100W
TMC100-S36
36V
2.78A
100W
TMC100-S48
48V
2.1A
100W
Notes: 1. Safety approvals are for PCB form only.
2. Add suffix "C" to order unit with cover fitted, e.g. TMC100-S18C, etc.
STANDARDS & COMPLIANCES
EN55011, EN55022 ........ Class B conducted and radiated
FCC .................................. Class B conducted and radiated
VCCI ................................. Class B conducted and radiated
EN 61000-3-2 .................. Harmonic distortion, Class A
EN 61000-3-3 .................. Line flicker
EN 61000-4-2 .................. ESD, ±8 KV air and ±6 KV contact
EN 61000-4-3 .................. Radiated immunity, 3V/m
EN 61000-4-4 .................. Fast transient/burst, ±2 KV
EN 61000-4-5 .................. Surge, ±1 KV diff., ±2 KV com.
EN 61000-4-6 .................. Conducted immunity, 3 Vrms
EN 61000-4-8 .................. Magnetic field immunity, 3A/m
EN 61000-4-11 ................ Voltage dip immunity, 30% reduction
for 500ms (criteria A @ 230 VAC,
criteria B @ 100 VAC), 60% reduction
for 100ms (criteria B), >95% reduction
for 10ms (criteria A)
Safety Standards ............. UL 60601-1, EN 60601-1,
CSA C22.2 No. 601.1,
UL 60950-1, EN 60950-1,
CSA C22.2 No. 60950-1
Agency Approvals ............ UL, cUL, TUV, CE, CB
Other Compliances ........ RoHS
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Building 13 • Santa Clara, California 95054, USA • 408-988-6616 • sales@trumpower.com • www.TRUMPower.com