TM66 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – 65 WATT MEDICAL POWER SUPPLIES | |||
|
TRUMPower
DESCRIPTION
The TM66 series of compact, open PCB constructed,
AC-DC switching power supplies are capable of
delivering 51 to 65 watts of continuous power. They
operate at 90 to 264VAC input voltage without the
need of voltage selection. They are ideally suited for
use in medical equipments. All models meet the
safety requirements of UL, CSA and IEC.
65 WATT MEDICAL POWER SUPPLIES
TM66 SERIES
( LV D )
RoHS
FEATURES
⧠Recognized or certified by UL, CSA and TÃV
⧠Small size
⧠100% burn-in
⧠90 to 264VAC wide input range
⧠Input surge current protection
⧠Overvoltage protection
⧠Overcurrent protection
⧠Compliant with RoHS requirement
INPUT SPECIFICATIONS
Input voltageï¼
90 to 264VAC
Input frequencyï¼
47 to 63 Hz
Input currentï¼
1.60Aï¼ rmsï¼ for 115VAC
1.00Aï¼ rmsï¼ for 230VAC
Earth leakage currentï¼ 90µA max. @ 115VAC, 60Hz
150µA max. @ 230VAC, 50Hz
OUTPUT SPECIFICATIONS
Output voltage/currentï¼ See rating chart
Total output powerï¼
65 watts maximum
Ripple and noiseï¼
1% peak to peak maximum
Overvoltage protectionï¼ Provided on output; set at
112 â 132% of its nominal output
voltage
Overcurrent protectionï¼ The output protected to short
circuit conditions
Temperature coefficientï¼ All outputs ±0.04%/ºC maximum
Transient responseï¼
Maximum excursion of 4%
or better on all models,
recovering to 1% of final value
within 500µs after a 25% step
load change
ENVIRONMENTAL SPECIFICATIONS
Operating temperatureï¼
Storage temperatureï¼ 0ºC to +70ºC
Relative humidityï¼
-40ºC to +85ºC
Deratingï¼
5% to 95% non-condensing
Derate from 100% at +50ºC linearly
to 50% at +70ºC
Safety Standard Approvalsï¼
UL 2601-1
CSA C22.2 No. 601.1
TÃV EN60601-1
GENERAL SPECIFICATIONS
Switching frequencyï¼ 42KHz ±5KHz
Efficiencyï¼
75% minimum on models with
Vo â¥12V, 6% minimum on the others
Hold-up timeï¼
10 msec minimum at 110VAC
Line regulationï¼
Inrush currentï¼
Withstand voltageï¼
±0.5% maximum at full load
17 amps @ 115VAC or 40 amps
@230VAC, at 25ºC cold start
4000VAC from input to output
1500VAC from input to ground
MTBFï¼
500VAC from output to ground
400,000 hours minimum at full load
at 25ºC ambient , calculated per
MIL-HDBK-217F
EMC Performance (EN60601-1-2: 2001)
EN55011:
FCC:
VCCI:
Class B conducted, Class B radiated
Class B conducted, Class B radiated
Class B conducted, Class B radiated
EN61000-3-2:
EN61000-3-3:
Harmonic distortion, Class A
Line flicker
EN61000-4-2:
ESD, ±8KV air and ±6KV contact
EN61000-4-3:
Radiated immunity, 3V/m for
EN61000-4-4:
80~2500MHz
Fast transient/burst, ±2KV
EN61000-4-5:
Surge, ±1KV diff, ±2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
EN61000-4-8:
EN61000-4-11:
Magnetic field immunity, 3A/m
Voltage dips, 30% reduction for 500ms,
60% reduction for 100ms and >95%
reduction for 10ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 â Fax: 408-988-6622 â email: sales@trumpower.com â Website: www.TRUMPower.com
|
▷ |