English
Language : 

TM650 Datasheet, PDF (1/3 Pages) Tumbler Technologies + TRUMPower – 650 WATT MEDICAL POWER SUPPLIES
TRUMPower
650 WATT MEDICAL POWER SUPPLIES
DESCRIPTION
The TM650 series comprising single and
multiple output models for 650-700 watts of
continuous output power is specially designed
for medical applications not for life support.
They operate at 90 to 264VAC input voltage
without the need of a selector strap. The units
are constructed on a printed circuit board with a
U-bracket for mechanical support and heat
sinking. A cover-and-fan assembly can be
added during manufacturing.
TM650 SERIES
(LVD)
RoHS
F E AT U R E S
‧ EN61000-3-2 class A and D compliant
‧ Power Factor 0.98 typical
‧ Overvoltage protection
‧ Short-circuit protection
‧ Thermal protection
‧ Power Fail Detect (PFD)
‧ 100% burn-in at full rated load
‧ Remote sense on output #1 and output #2
‧ Remote inhibit – TTL high disables output
‧ Compliant with RoHS requirements
Safety Standard Approvals:
UL 60601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
INPUT SPECIFICATIONS
Input voltage:
Input frequency:
90 to 264VAC
47 to 63Hz
Input current:
Earth leakage current:
10A(rms)for 115VAC
5A(rms)for 230VAC
100µA max. @ 115VAC, 60Hz
165µA max. @ 230VAC, 50H z
ENVIRONMENTAL SPECIFICATIONS
Operating temperature: 0ºC to +70ºC
Storage temperature: -40ºC to +85ºC
Relative humidity:
5% to 95% non-condensing
Derating:
Derate from 100% at +50ºC
linearly to 50% at +70ºC
OUTPUT SPECIFICATIONS
Output voltage/current: See rating chart
Total output power:
See rating chart
Ripple and noise:
2% peak to peak maximum
Overvoltage protection: Provided on output #1 only; set
at 115 –140% of its nominal output
voltage
Overcurrent protection: All outputs protected to short
circuit conditions
Temperature coefficient: All outputs ±0.04%/ºC maximum
Transient response:
Maximum excursion of 4% or
better on all models, recovering to
1% of final value within 500µs
after a 25% step load change
PFD signal:
TTL logic high for normal
operation and TTL logic low upon
loss of input power. This signal
appears at least 1ms prior to
master output dropping 5% below
its nominal value. This signal also
provides a minimum delay of 100
ms after master output is within
regulation.
Remote inhibit:
Requires an external TTL high
level signal to inhibit outputs for
standard models.
GENERAL SPECIFICATIONS
Switching frequency:
Power factor:
Efficiency:
Hold-up time:
70KHz ±10KHz
0.98 typical
80% minimum on all models
12 msec minimum at 110VAC
Line regulation:
Inrush current:
±0.2% maximum at full load
50amps @115VAC or 100 amps
@ 230VAC at 25ºC cold start
Withstand voltage: 4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
MTBF:
300,000 hours minimum at full load
at 25ºC ambient, calculated per
MIL-HDBK- 217F
EMC Performance (EN60601-1-2)
EN55011:
EN61000-3-2:
EN61000-3-3:
Class B conducted, Class A radiated
Harmonic distortion, Class A and D
Line flicker
EN61000-4-2:
ESD, ±8KV air and ±6KV contact
EN61000-4-3:
Radiated immunity, 3V/m
EN61000-4-4:
Fast transient/burst, ±2KV
EN61000-4-5:
EN61000-4-6:
Surge, ±1KV diff, ±2KV com.
Conducted immunity, 3Vrms
EN61000-4-8:
Magnetic field immunity, 3A/m
EN61000-4-11:
Voltage dips, 30% reduction for
500ms, 60% reduction for 100ms
and >95% reduction for 10ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA