English
Language : 

TM25 Datasheet, PDF (1/3 Pages) Tumbler Technologies + TRUMPower – 25-28 WATT MEDICAL POWER SUPPLIES
TRUMPower 25-28 WATT MEDICAL POWER SUPPLIES
DESCRIPTION
The TM25 series of compact, open PCB constructed,
AC-DC switching power supplies are capable of
delivering 25 to 28 watts of continuous output
power. They operate at 85 to 264VAC input voltage
without the need of voltage selection. They are
ideally suited for use in medical equipments, safety
systems and monitoring equipments.
TM25 SERIES
( LV D )
RoHS
FEATURES
‧ Low safety ground leakage current
‧ 100% burn-in
‧ 85 to 264VAC wide input range
‧ Input surge current protection
‧ Overvoltage protection
‧ Overcurrent protection
‧ Open PCB construction
‧ Compliant with RoHS requirements
INPUT SPECIFICATIONS
Input voltage:
85 to 264VAC
Input frequency:
47 to 63 Hz
Input current:
0.7A( rms) for 115VAC
0.4A( rms) for 230VAC
E a r t h l e a k a g e c u r r e n t: 9 0 µ A m a x . @ 11 5 VA C , 6 0 H z
150µA max. @ 230VAC, 50Hz
OUTPUT SPECIFICATIONS
Output voltage/current: See Rating Chart
Total output power:
See Rating Chart
Ripple and noise:
1% peak to peak max.
Overvoltage protection: Provided on output #1 only;
set at 112-132% of its
nominal output voltage
Overcurrent protection: All outputs protected to short
circuit conditions
Temperature coefficient: All outputs ±0.04%/ºC
maximum
Transient response:
Maximum excursion of 4%
or better on all models,
recovering to 1% of final
value within 500µs after a
25% step load change
ENVIRONMENTAL SPECIFICATIONS
Operating temperature: 0ºC to +70ºC
Storage temperature: -40ºC to +85ºC
Relative humidity:
5% to 95% non-condensing
Derating:
Derate from 100% at +50ºC
linearly to 50% at +70ºC
Safety Standard Approvals:
UL 2601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
GENERAL SPECIFICATIONS
Switching frequency: 42KHz ±5KHz
Efficiency:
70% minimum on single output
model with Vo ≥ 12V, 68% minimum
on the others
Hold-up time:
10 msec minimum at 110VAC
Line regulation:
±0.5% maximum at full load
Inrush current:
10 amps @ 115VAC or 25 amps
@ 230VAC at 25ºC cold start
Withstand voltage:
4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
MTBF:
500,000 hours minimum at full load
at 25ºC ambient, calculated per
MIL- HDBK- 217F
EMC Performance (EN60601-1-2:2001)
EN55011:
FCC:
VCCI:
Class B conducted, Class B radiated
Class B conducted, Class B radiated
Class B conducted, Class B radiated
EN61000-3-2: Harmonic distortion, Class A and D
EN61000-3-3: Line flicker
EN61000-4-2: ESD, ±8KV air and ±6KV contact
E N 6 1 0 0 0 - 4 - 3 : Radiated immunity, 3V/m for 80~2500 MHz
EN61000-4-4: Fast transient/burst, ±2KV
EN61000-4-5: Surge, ±1KV diff, ±2KV com.
E N 6 1 0 0 0 - 4 - 6 : Conducted immunity, 3Vrms
E N 6 1 0 0 0 - 4 - 8 : Magnetic field immunity, 3A/m
EN61000-4-11:
Voltage dips, 30% reduction for 500ms,
60% reduction for 100ms and >95%
reduction for 10ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: sales@trumpower.com ● Website: www.TRUMPower.com