TRUMPower 150- 200 WATT MEDICAL POWER SUPPLIES
DESCRIPTION
The TM201 series comprising single and multiple
output models for 150 to 200 watts of continuous
output power is specially designed for dental and
medical applications. They operate at 90 to
264VAC input voltage without the need of a
selector strap. All auxiliary outputs are with
magnetic amplifier linear regulator to keep
regulation. The units are constructed on a printed
circuit board with a U-bracket for mechanical
support and heat sinking. A cover-and-fan
assembly can be added during manufacturing.
TM201 SERIES
( LV D )
RoHS
F E AT U R E S
‧ Low safety ground leakage current
‧ Meet EN 55011 and FCC Class B
‧ Power Factor 0.98 typical
‧ Short-circuit protection
‧ Power Fail Detect( PFD) signal
‧ 100% burn-in at full rated load
‧ Optional cover-and-fan assembly
‧ Remote inhibit - TTL high disables output
‧ Compliant with RoHS requirements
INPUT SPECIFICATIONS
Input voltage:
Input frequency:
90 to 264VAC
47 to 63Hz
Input current:
3.2A( rms) for 115VAC
1.6A( rms) for 230VAC
Earth leakage current: 90µA max. @ 115VAC, 60Hz
150µA max. @ 230VAC, 50Hz
OUTPUT SPECIFICATIONS
Output voltage/current:
Total output power:
Ripple and noise:
Overvoltage protection:
Overcurrent protection:
Temperature coefficient:
Transient response:
PFD signal:
See rating chart
200 watts maximum
2% peak to peak maximum
Provided on output #1 only;
set at 112-132% of its nominal
output voltage
All outputs protected to short
circuit conditions
All outputs ±0.04/ºC maximum
Maximum excursion of 4% or
better on all models,
recovering to 1% of final
value within 500µs after a
25% step load change
TTL logic high for normal
operation and TTL logic low
upon loss of input power. This
signal appears at least 1ms
prior to +5V output dropping
5% below its nominal value.
This signal also provides a
minimum delay of 100ms after
+5V is within regulation.
Safety Standard Approvals:
UL 60601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
ENVIRONMENTAL SPECIFICATIONS
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
Cooling:
0ºC to +70ºC
-40ºC to +85ºC
5% to 95% non-condensing
Derate from 100% at +50ºC
linearly to 50% at +70ºC
10.8 CFM forced air provided on “C”
version; 25 CFM forced air to be
provided for U-bracket version by
user.
GENERAL SPECIFICATIONS
Switching frequency: 88 KHz –112 KHz
Efficiency:
70% minimum on all models
Hold-up time:
20 msec minimum at 110VAC
Line regulation:
±0.5% maximum at full load
Inrush current:
20 amps @ 115VAC or 40 amps
@ 230VAC, at 25ºC cold start
Withstand voltage: 4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
MTBF:
350,000 hours at full load at
25ºC ambient, calculated per
MIL-HDBK-217F
EMC Performance (EN60601-1-2: 2001)
EN55011:
Class B conducted, Class B
radiated
EN61000-3-2:
Harmonic distortion, Class A and D
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD, ±8KV air and ±6KV contact
EN61000-4-3:
Radiated immunity, 3V/m
EN61000-4-4:
Fast transient/burst, ±2KV
EN61000-4-5:
Surge, ±1KV diff, ±2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
EN61000-4-8:
Magnetic field immunity, 3A/m
EN61000-4-11:
Voltage dips, 30% reduction for
500ms, 60% reduction for 100ms
and >95% reduction for 10ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: sales@trumpower.com ● Website: www.TRUMPower.com