TM200 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – 200 WATT MEDICAL POWER SUPPLIES
TRUMPower
200 WATT MEDICAL POWER SUPPLIES
DESCRIPTION
Comprising single and multiple output models
with 200 watts of continuous output power, the
TM200 series is specially designed for medical
applications. Auto-ranging input allows them to
operate from any line voltage throughout the
world without a selector strap. All auxiliary
outputs are with magnetic amplifier or linear
regulator to keep regulation. The units are
constructed on a printed circuit board with a
U-bracket for mechanical support and heat
sinking. A cover-and-fan assembly can be
added during manufacturing.
F E AT U R E S
‧ Low safety ground leakage current
‧ Meet EN 55011 and FCC class B
‧ Auto-ranging input
‧ Short-circuit protection with auto-recovery
‧ Power Fail Detect( PFD) signal
‧ 100% burn-in at full rated load
‧ Optional cover-and-fan assembly
INPUT SPECIFICATIONS
Input voltage:
Input frequency:
Input current:
Leakage current:
90 to 130 VAC, or 180 to
264 VAC( auto-ranging)
47 to 63 Hz
4.5A( rms) for 115VAC
2.6A( rms) for 230VAC
90 µA max. @ 115VAC, 60Hz
15 µA max. @ 230VAC, 50Hz
OUTPUT SPECIFICATIONS
Output voltage/current: See rating chart
Total output power: 200 watts maximum
Ripple and noise:
2% peak to peak maximum
Overvoltage protection: Provided on output #1 only; set at
112 – 132% of its nominal output
voltage
Overcurrent protection: All outputs protected to short
circuit conditions
Temperature coefficient:All outputs ±0.04%/ºC maximum
Transient response:
Maximum excursion of 4% or
better on all models, recovering to
1% of final value within 500µs
after a 25% step load change
PFD signal:
TTL logic high for normal
operation and TTL logic low upon
loss of input power. This signal
appears at least 1 ms prior to +5V
output dropping 5% below its
nominal value. This signal also
provides a minimum delay of 100
ms after +5V is within regulation.
TM200 SERIES
(MDD)
Safety Standard Approvals:
UL 2601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
ENVIRONMENTAL SPECIFICATIONS
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
0ºC to +70ºC
-40ºC to +85ºC
5% to 95% non-condensing
Derate from 100% at +50ºC
linearly to 50% at +70ºC
Cooling:
25 CFM forced air, provided
on covered version, to be
provided for U-bracket
version by user
GENERAL SPECIFICATIONS
Switching frequency: 64KHz ±5KHz
Efficiency:
70% minimum on all models
Hold-up time:
12 msec minimum at 110VAC
Line regulation:
±0.5% maximum at full load
Inrush current:
20 amps @ 115VAC, or 40 amps @
230VAC, at 25ºC cold start
Withstand voltage: 4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
MTBF:
450,000 hours at full load at 25ºC
ambient, calculated per
MIL-HDBK-217F
EMC Performance (EN60601-1-2: 2001)
EN55011:
Class B conducted, Class B radiated
FCC:
Class B conducted, Class B radiated
VCCI:
Class B conducted, Class B radiated
EN61000-3-2:
Harmonic distortion, Class A
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD, ±8KV air and ±4KV contact
EN61000-4-3:
Radiated immunity, 3V/m
EN61000-4-4:
Fast transient/burst, ±1KV
EN61000-4-5:
Surge, ±1KV diff, ±2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
EN61000-4-8:
Magnetic field immunity, 1A/m
EN61000-4-11:
Voltage dips, 30% reduction for 500ms
and >95% reduction for 10ms
Interruptions, >95% for 5000ms