TM200 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – 200 WATT MEDICAL POWER SUPPLIES | |||
|
TRUMPower
200 WATT MEDICAL POWER SUPPLIES
DESCRIPTION
Comprising single and multiple output models
with 200 watts of continuous output power, the
TM200 series is specially designed for medical
applications. Auto-ranging input allows them to
operate from any line voltage throughout the
world without a selector strap. All auxiliary
outputs are with magnetic amplifier or linear
regulator to keep regulation. The units are
constructed on a printed circuit board with a
U-bracket for mechanical support and heat
sinking. A cover-and-fan assembly can be
added during manufacturing.
F E AT U R E S
⧠Low safety ground leakage current
⧠Meet EN 55011 and FCC class B
⧠Auto-ranging input
⧠Short-circuit protection with auto-recovery
⧠Power Fail Detectï¼ PFDï¼ signal
⧠100% burn-in at full rated load
⧠Optional cover-and-fan assembly
INPUT SPECIFICATIONS
Input voltage:
Input frequency:
Input current:
Leakage current:
90 to 130 VAC, or 180 to
264 VACï¼ auto-rangingï¼
47 to 63 Hz
4.5Aï¼ rmsï¼ for 115VAC
2.6Aï¼ rmsï¼ for 230VAC
90 µA max. @ 115VAC, 60Hz
15 µA max. @ 230VAC, 50Hz
OUTPUT SPECIFICATIONS
Output voltage/currentï¼ See rating chart
Total output powerï¼ 200 watts maximum
Ripple and noiseï¼
2% peak to peak maximum
Overvoltage protectionï¼ Provided on output #1 only; set at
112 â 132% of its nominal output
voltage
Overcurrent protectionï¼ All outputs protected to short
circuit conditions
Temperature coefficientï¼All outputs ±0.04%/ºC maximum
Transient responseï¼
Maximum excursion of 4% or
better on all models, recovering to
1% of final value within 500µs
after a 25% step load change
PFD signalï¼
TTL logic high for normal
operation and TTL logic low upon
loss of input power. This signal
appears at least 1 ms prior to +5V
output dropping 5% below its
nominal value. This signal also
provides a minimum delay of 100
ms after +5V is within regulation.
TM200 SERIES
(MDD)
Safety Standard Approvals:
UL 2601-1
CSA C22.2 No. 601.1
TÃV EN60601-1
ENVIRONMENTAL SPECIFICATIONS
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
0ºC to +70ºC
-40ºC to +85ºC
5% to 95% non-condensing
Derate from 100% at +50ºC
linearly to 50% at +70ºC
Cooling:
25 CFM forced air, provided
on covered version, to be
provided for U-bracket
version by user
GENERAL SPECIFICATIONS
Switching frequencyï¼ 64KHz ±5KHz
Efficiencyï¼
70% minimum on all models
Hold-up timeï¼
12 msec minimum at 110VAC
Line regulationï¼
±0.5% maximum at full load
Inrush currentï¼
20 amps @ 115VAC, or 40 amps @
230VAC, at 25ºC cold start
Withstand voltageï¼ 4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
MTBFï¼
450,000 hours at full load at 25ºC
ambient, calculated per
MIL-HDBK-217F
EMC Performance (EN60601-1-2: 2001)
EN55011:
Class B conducted, Class B radiated
FCC:
Class B conducted, Class B radiated
VCCI:
Class B conducted, Class B radiated
EN61000-3-2:
Harmonic distortion, Class A
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD, ±8KV air and ±4KV contact
EN61000-4-3:
Radiated immunity, 3V/m
EN61000-4-4:
Fast transient/burst, ±1KV
EN61000-4-5:
Surge, ±1KV diff, ±2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
EN61000-4-8:
Magnetic field immunity, 1A/m
EN61000-4-11:
Voltage dips, 30% reduction for 500ms
and >95% reduction for 10ms
Interruptions, >95% for 5000ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 â Fax: 408-988-6622 â email: sales@trumpower.com â Website: www.TRUMPower.com
|
▷ |