English
Language : 

TM1100 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – 1100 Watts AC/DC Medical & ITE Grade Power Supplies
TRUMPower
TM1100 Series
1100 Watts AC/DC Medical & ITE Grade Power Supplies
FEATURES
• BF Rated
• 5,000 Meters Operating Altitude
• 5.91" x 9.25" x 2.40" Compact Size
• Low Leakage Current
• Efficiency up to 89%
INPUT SPECIFICATIONS
Input Voltage Range .............. 90-264 VAC, with active PFC
Input Frequency ..................... 47-63Hz
Input Current .......................... 16A rms @ 100 VAC, 60Hz
8A rms @ 240 VAC, 50Hz
Inrush Current ....................... 50A @ 115 VAC or 100A @ 230
............................................... VAC, at 25°C cold start
Earth Leakage Current .......... 220µA max. @ 264 VAC, 63Hz
OUTPUT SPECIFICATIONS
Output Power Ratings ........... See table
Output Voltage ....................... See table
Tolerance ............................... ±2%
Ripple and Noise* ................. 1% peak to peak max.
Remote Sense ...................... Compensation for cable losses up
to 0.5V
Over Voltage Protection ......... Set at 112-140% of its nominal
output voltage
Over Current Protection ........ Set at 120-140% of maximum
output current
Thermal Shutdown ................ Protected to over-temperature
conditions
Temperature Coefficient ....... ±0.04%/°C max.
Transient Response ............. Max. excursion of 4% or better on
all models, recovering to 1% of
final value within 500 µs after a
25% step load change
Standby Power ...................... 5V/200mA max.
Fan Power ............................. 12V/1A max.
* Peak to peak with 20MHz bandwidth and 10µF tantalum capacitor in parallel with a
0.1µF ceramic capacitor at rated line voltage and load ranges
MODELS LIST
Product No.
Voltage
Max.
Current
Output
Max.
Power
Peak
Peak Ripple &
Current (1) Power (1) Noise*
TM1100-S24C
24V
45.84A
1100W
52.10A
1250W 240mV
TM1100-S28C
28V
39.29A 1100W
44.65A
1250W 280mV
TM1100-S32C
32V
34.38A 1100W
39.07A
1250W 320mV
TM1100-S34C
34V
32.25A 1100W
36.77A
1250W 340mV
TM1100-S36C
36V
30.56A 1100W
34.73A
1250W 360mV
TM1100-S42C
42V
26.20A 1100W
29.77A
1250W 420mV
TM1100-S48C
48V
22.92A 1100W
26.10A
1250W 480mV
NOTES:
1. Peak current and power possible at 170-260 VAC input, 10 seconds, 35% duty
cycle.
GENERAL SPECIFICATIONS
Switching Frequency ............. 40 KHz-200 KHz
Power Factor ......................... 0.98 typical
Efficiency ............................... 87% typical @ 115 VAC,
89% typical @ 230 VAC
Operating Altitude .................. 5,000 meters max.
Hold-up Time ......................... 10ms @ 110 VAC
Line Regulation ..................... ±0.5% max. @ full load
Load Regulation ................... ±2% max. @ full load
Operating Temperature ........ -10°C to +70°C
Storage Temperature ............ -40°C to +85°C
Derating ................................. Derate from 100% @ +50°C
linearly to 50% @ +70°C,
applicable to convection and
forced-air cooling conditions
Relative Humidity .................. 5% to 95% non-condensing
Withstand Voltage ................. 4,000 VAC input-output (2MOPP)
............................................... 1,500 VAC input-ground (1MOPP)
............................................... 1,500 VAC output-ground (1MOPP)
MTBF ...................................... 300K hours minimum at full load,
25°C ambient, calculated per
MIL-HDBK-217F
STANDARDS & COMPLIANCES
IEC 60601-1-2 ................... EMC Performance
EN55011 ........................... Class B conducted, Class B radiated
FCC, VCCI ......................... Class B conducted, Class B radiated
EN61000-3-2 ..................... Harmonic distortion, Class A & D
EN61000-3-3 ..................... Line flicker
EN61000-4-2 ..................... ESD, ±8 KV air and ±6 KV contact
EN61000-4-3 ..................... Radiated immunity, 3V/m
EN61000-4-4 ..................... Fast transient/burst, ±2 KV
EN61000-4-5 ..................... Surge, ±1 KV diff., ±2 KV com.
EN61000-4-6 ..................... Conducted immunity, 3 Vrms
EN61000-4-8 ..................... Magnetic field immunity, 3A/m
EN61000-4-11 ................... Voltage dip immunity,
30% reduction for 500ms,
60% reduction for 100ms,
>95% reduction for 10ms
Safety Standards ............... UL/IEC/EN 60601-1 (3rd Edition)
ANSI/AAMI ES 60601-1:2005
CSA C22.2 No. 60601-1:2008
UL/IEC/EN 60950-1 (2nd Edition)
Agency Approvals .............. UL, TUV, CB, CE
Other Compliance ............. RoHS
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Building 13 • Santa Clara, California 95054, USA • 408-988-6616 • sales@trumpower.com • www.TRUMPower.com