English
Language : 

THM30 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – Medical Grade 30W Open Frame Power Supplies
TRUMPower
FEATURES
• 1.5" x 4" Footprint
• Class II with 2 MOPP
• 5V to 40V Output Voltages
• 3,000 Meters Operating Altitude
• 0.5W No Load Power Consumption
• UL, cUL, TUV, FCC, CB & CE Approvals
THM30 Series
Medical Grade 30W Open Frame Power Supplies
INPUT SPECIFICATIONS
Input Voltage Range ................. 80-275 VAC
Input Frequency ........................ 47-63 Hz
Input Current ............................. 0.8A rms max. @ 100 VAC,
0.4A rms max. @ 240 VAC
Inrush Current .......................... 25A @ 100 VAC or 50A
@ 240 VAC, at 25°C cold start
No Load Power Consumption... 0.5W max.
OUTPUT SPECIFICATIONS
Output Power Ratings .............. See models list
Minimum Load ......................... Not required
Line Regulation ........................ ±1% max. @ full load
Total Regulation ....................... See models list
Ripple & Noise ......................... See models list
Overload Protection ................. Set at 110-150% of its nominal
output voltage; Auto recovery
Short Circuit Protection ............ Auto recovery
Temperature Coefficient .......... ±0.04%/°C max.
Transient Response ................ 4ms @ 110 VAC, full load
GENERAL SPECIFICATIONS
Efficiency .................................. See models list
Switching Frequency ................ 65 kHz typ.
Power Factor ............................ 0.65 typ.
Hold-up Time ............................ 12ms max. @ 100 VAC, full load
Start-up Time ............................ 0.5 secs. max. @ full load
Operating Temperature ........... -10°C to +70°C
Storage Temperature ............... -40°C to +85°C
Operating Humidity .................. 0% to 95% non-condensing
Storage Humidity ...................... 0% to 95% non-condensing
Derating .................................... Derate from 100% at +40ºC
linearly to 50% at +70°C
Withstand Voltage .................... 4,000 VAC input-output
Operating Altitude ..................... 3,000 meters max.
MTBF ......................................... 100 kHrs minimum at full load,
25°C ambient, calculated per
MIL-HDBK-217F
STANDARDS & COMPLIANCES
IEC/EN 60601-1-2 ............... EMC & Immunity Performance
EN 55011, CISPR11, FCC .. Class B, conducted & radiated
EN 61000-3-2 ...................... Harmonic distortion, Class B & D
EN 61000-3-3 ...................... Line flicker
EN 61000-4-2 ...................... ESD, ±15 KV air and ±8 KV contact
EN 61000-4-3 ...................... Radiated immunity, 3V/m
EN 61000-4-4 ...................... Fast transient/burst, ±2 KV
EN 61000-4-5 ...................... Surge, ±1 KV diff., ±2 KV com.
EN 61000-4-6 ...................... Conducted immunity, 3 Vrms
EN 61000-4-8 ...................... Magnetic field immunity, 3A/m
EN 61000-4-11 .................... Voltage dips,
30% reduction for 500ms,
60% reduction for 100ms,
>95% reduction for 10ms
Safety Standards ................. IEC/EN 60601-1 3rd Edition,
............................................. ANSI/AAMI ES 60601-1:2005,
............................................. CSA C22.2 No. 60601-1:14
Agency Approvals ................ UL, cUL, TUV, FCC, CE, CB
Other Compliances ............ RoHS2
MECHANICAL SPECIFICATIONS
OUTPUT CONNECTOR (P2)
PIN
FUNCTION
1
+V
2
+V
3
RETURN
4
RETURN
Notes:
1. Unit: inch [mm]
2. Tolerance: 0.02 [0.5] max.
3. Weight: 85g (0.19 lbs) approx.
4. Input Connector: Mates with Molex housing
09-50-3031 and Molex 2478 series crimp
terminal.
5. Output Connector: Mates with Molex
housing 09-50-3041 and Molex 2478 series
crimp terminal.
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Building 13 • Santa Clara, California 95054, USA • 408-988-6616 • sales@trumpower.com • www.TRUMPower.com