English
Language : 

TAM30 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – 30 Watts Medical Grade AC/DC Power Adapters
TRUMPower
TAM30 Series
30 Watts Medical Grade AC/DC Power Adapters
FEATURES
• 90~264 VAC Universal Input
• Class I with IEC 320/C14 AC Inlet
• Power-on LED Indicator
• Low Earth Leakage Current
• Standby Consumption < 0.3W
• Meet CEC & Energy Star Efficiency Level V Requirements
MODELS LIST
Product No.
Output
Voltage
Min. Output
Current
Max. Output
Current
Tolerance
Ripple & Noise
(mVp-p)*
Average Efficiency (Typical)
@ 115 VAC
@ 230 VAC
Max. Output
Power
TAM30-05
5V
0A
5.0A
±5%
75mV
83%
81%
25W
TAM30-09
9V
0A
3.33A
±5%
90mV
86%
84%
30W
TAM30-12
12V
0A
2.5A
±5%
120mV
86%
85%
30W
TAM30-15
15V
0A
2.0A
±5%
150mV
88%
86%
30W
TAM30-18
18V
0A
1.66A
±5%
180mV
88%
86%
30W
TAM30-24
24V
0A
1.25A
±5%
240mV
89%
87%
30W
TAM30-48
48V
0A
0.625A
±5%
480mV
91%
90%
30W
* Measured at rated line voltage and output load ranges, 20 MHz BW, with a 10µF tantalum capacitor in parallel with a 0.1µF ceramic capacitor across the output.
INPUT SPECIFICATIONS
Input Voltage Range ............ 90~264 VAC
Input Frequency ................... 47~63Hz
Input Current ........................ 1.0A (rms) @ 115 VAC
0.6A (rms) @ 230 VAC
Inrush Current ..................... 50A @ 115 VAC or 100A @ 230 VAC,
at 25°C cold start
Earth Leakage Current ........ 100µA max. @ 264 VAC, 63Hz
GENERAL SPECIFICATIONS
Efficiency ............................. See table
Hold-up Time ....................... 8 ms min. at 115 VAC
Turn on Delay Time ............. 3 seconds max. at 115 VAC
Switching Frequency ........... 67 KHz
Line Regulation ................... ±0.5% max. at full load
Operating Temperature ...... 0°C to +50°C
Derating ............................... Derate from 100% at +40°C linearly
to 50% at +60°C
Storage Temperature .......... -40°C to +85°C
Relative Humidity ................ 10% to 90% non-condensing
Withstand Voltage ............... 4,000 VAC from input to output
MTBF .................................... 300K hrs min. at full load,
25°C ambient, calculated per
MIL-HDBK-217F
OUTPUT SPECIFICATIONS
Output Power Ratings ......... See table
Ripple and Noise ................ See table
Overvoltage Protection ........ Set at 116-230% of its nominal
output voltage
Overcurrent Protection ........ Protected to short-circuit conditions
Transient Response ........... Max. excursion of 4% or better on
all models, recovering to 1% of
final value within 500 µs after a
25% step load change
STANDARDS & COMPLIANCES
IEC 60601-1-2 ..................... EMC performance
EN 55011, FCC, VCCI ......... Class B, conducted & radiated
EN 61000-3-2 ...................... Harmonic distortion, Class A
EN 61000-3-3 ...................... Line flicker
EN 61000-4-2 ...................... ESD, ±8 KV air and ±6 KV contact
EN 61000-4-3 ...................... Radiated immunity, 3V/m
EN 61000-4-4 ...................... Fast transient/burst, ±2 KV
EN 61000-4-5 ...................... Surge, ±1 KV diff., ±2 KV com.
EN 61000-4-6 ...................... Conducted immunity, 3 Vrms
EN 61000-4-8 ...................... Magnetic field immunity, 3A/m
EN 61000-4-11 .................... Voltage dip immunity,
30% reduction for 500ms,
60% reduction for 100ms,
>95% reduction for 10ms
Safety Standards ................. UL60601-1, EN 60601-1
CSA C22.2 No. 601.1
Agency Approvals ................ UL, cUL, TUV, CB, CE (LVD)
Other Compliances ............ RoHS, CEC & Energy Star Level V
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Building 13 • Santa Clara, California 95054, USA • 408-988-6616 • sales@trumpower.com • www.TRUMPower.com