English
Language : 

PM65 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – 65 WATT MEDICAL POWER SUPPLIES
65 WATT MEDICAL POWER SUPPLIES
DESCRIPTION
The PM65 series of compact, open PCB constructed,
AC-DC switching power supplies are capable of delivering
50 to 65 watts of continuous power. They are ideally
suited for use in medical equipment not for patient
contact, and operate at 85 to 264 VAC input voltage
without the need of voltage selection. All models meet the
safety requirements of UL, CSA and IEC.
PM65 SERIES
(MDD)
FEATURES
‧ Low safety ground leakage current
‧ Meet EN 55011 and FCC Class B
‧ Small size, light weight
‧ 100% burn-in
‧ W i d e i n p u t r a n g e 8 5 t o 2 6 4 VA C
‧ Input surge current protection
‧ Overvoltage protection
‧ Overcurrent protection
INPUT SPECIFICATIONS
Input voltage:
Input frequency:
Input current:
Leakage current:
85 to 264 VAC
47 to 63 Hz
1.60A(rms)for 115VAC
1.00A(rms)for 230VAC
90uA max. @ 115VAC, 60Hz
150uA max. @ 230VAC, 50Hz
Safety Standard Approvals:
UL 2601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
OUTPUT SPECIFICATIONS
Output voltage/current:
Total output power:
Ripple and noise:
Overvoltage protection:
Overcurrent protection:
Temperature coefficient:
Transient response:
See rating chart
65 watts maximum
1﹪peak to peak maximum
Provided on output #1 only; set at
112 - 132﹪of its nominal output
voltage
All outputs protected to short
circuit conditions
All outputs ±0.04﹪/℃maximum
Maximum excursion of 4﹪or better
on all models, recovering to 1﹪of
final value within 500us after a 25﹪
step load change
ENVIRONMENTAL SPECIFICATIONS
Operating temperature: 0℃ to +70℃
Storage temperature: -40℃ to +85℃
Relative humidity:
5﹪to 95﹪non-condensing
Derating:
Derate from 100﹪at +50℃ linearly
to 50% at +70℃
GENERAL SPECIFICATIONS
Switching frequency: 32KHz ±5KHz
Efficiency:
70﹪minimum on single output models with Vo
≧12V, 65% minimum on the others.
Hold-up time:
10 msec minimum at 110VAC
Line regulation:
±0.5﹪maximum at full load
Inrush current:
15 amps @ 115VAC, or 30 amps @ 230VAC, at
25℃ cold start
Withstand voltage: 4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
MTBF:
400,000 hours minimum at full load at 25℃
ambient , calculated per MIL-HDBK-217F
EMC PERFORMANCE (EN60601-1-2: 2001)
EN55011:
Class B conducted, Class B radiated
FCC:
Class B conducted, Class B radiated
VCCI:
Class B conducted, Class B radiated
EN61000-3-2:
Harmonic distortion , Class A
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD,± 8KV air and ± 6KV contact
EN61000-4-3:
Radiated immunity, 3V/m, for 80~2500MHz
EN61000-4-4:
Fast transient/burst, ± 2KV
EN61000-4-5:
Surge, ± 1KV diff., ± 2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
EN61000-4-8:
Magnetic field immunity, 3A/m
EN61000-4-11:
Voltage dips, 30% reduction for 500ms, 60%
reduction for 100ms and >95% reduction for
10ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 ● sales@trumpower.com ● www.TRUMPower.com