PM65 Datasheet, PDF (1/2 Pages) Tumbler Technologies + TRUMPower – 65 WATT MEDICAL POWER SUPPLIES | |||
|
65 WATT MEDICAL POWER SUPPLIES
DESCRIPTION
The PM65 series of compact, open PCB constructed,
AC-DC switching power supplies are capable of delivering
50 to 65 watts of continuous power. They are ideally
suited for use in medical equipment not for patient
contact, and operate at 85 to 264 VAC input voltage
without the need of voltage selection. All models meet the
safety requirements of UL, CSA and IEC.
PM65 SERIES
(MDD)
FEATURES
⧠Low safety ground leakage current
⧠Meet EN 55011 and FCC Class B
⧠Small size, light weight
⧠100% burn-in
⧠W i d e i n p u t r a n g e 8 5 t o 2 6 4 VA C
⧠Input surge current protection
⧠Overvoltage protection
⧠Overcurrent protection
INPUT SPECIFICATIONS
Input voltageï¼
Input frequencyï¼
Input currentï¼
Leakage currentï¼
85 to 264 VAC
47 to 63 Hz
1.60Aï¼rmsï¼for 115VAC
1.00Aï¼rmsï¼for 230VAC
90uA max. @ 115VAC, 60Hz
150uA max. @ 230VAC, 50Hz
Safety Standard Approvalsï¼
UL 2601-1
CSA C22.2 No. 601.1
TÃV EN60601-1
OUTPUT SPECIFICATIONS
Output voltage/currentï¼
Total output powerï¼
Ripple and noiseï¼
Overvoltage protectionï¼
Overcurrent protectionï¼
Temperature coefficientï¼
Transient responseï¼
See rating chart
65 watts maximum
1﹪peak to peak maximum
Provided on output #1 only; set at
112 - 132﹪of its nominal output
voltage
All outputs protected to short
circuit conditions
All outputs ±0.04﹪/âmaximum
Maximum excursion of 4﹪or better
on all models, recovering to 1﹪of
final value within 500us after a 25﹪
step load change
ENVIRONMENTAL SPECIFICATIONS
Operating temperatureï¼ 0â to +70â
Storage temperatureï¼ -40â to +85â
Relative humidityï¼
5﹪to 95﹪non-condensing
Deratingï¼
Derate from 100﹪at +50â linearly
to 50% at +70â
GENERAL SPECIFICATIONS
Switching frequencyï¼ 32KHz ±5KHz
Efficiencyï¼
70﹪minimum on single output models with Vo
â§12V, 65% minimum on the others.
Hold-up timeï¼
10 msec minimum at 110VAC
Line regulationï¼
±0.5﹪maximum at full load
Inrush currentï¼
15 amps @ 115VAC, or 30 amps @ 230VAC, at
25â cold start
Withstand voltageï¼ 4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
MTBFï¼
400,000 hours minimum at full load at 25â
ambient , calculated per MIL-HDBK-217F
EMC PERFORMANCE (EN60601-1-2: 2001)
EN55011:
Class B conducted, Class B radiated
FCC:
Class B conducted, Class B radiated
VCCI:
Class B conducted, Class B radiated
EN61000-3-2:
Harmonic distortion , Class A
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD,± 8KV air and ± 6KV contact
EN61000-4-3:
Radiated immunity, 3V/m, for 80~2500MHz
EN61000-4-4:
Fast transient/burst, ± 2KV
EN61000-4-5:
Surge, ± 1KV diff., ± 2KV com.
EN61000-4-6:
Conducted immunity, 3Vrms
EN61000-4-8:
Magnetic field immunity, 3A/m
EN61000-4-11:
Voltage dips, 30% reduction for 500ms, 60%
reduction for 100ms and >95% reduction for
10ms
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616 â sales@trumpower.com â www.TRUMPower.com
|
▷ |