English
Language : 

TPMPU12C Datasheet, PDF (1/2 Pages) Total Power International – 12W Wall Moumt Switching Power Supplies For Medical Equipment
TPMPU12C SERIES
12W Wall Mount Switching Power Supplies For Medical Equipment
Description:
The 12C series of AC/DC switching mode power supplies
provide 12 watts of continuous output power.
This series is suitable for use in Blood Pressure measurements,
Frequency Therapy Device and Dental Curing light applications.
All models are designed to comply with TUV/T-mark (EN 60601-
1) and new CE requirements.
Features:
Wide Input Voltage 90 to 264 VAC,47 to 63 Hz
2 Prong Plug-In Mains Connector
Output Voltage Available From 5VDC Thru 27VDC
Optional Output Connector (See appendix)
Single Output
Class II
Over Voltage and Over Load protection.
Over temperature Detection
3 year warranty
Electrical Characteristics:
Safety Approvals:
Sym.
Parameter
Test Conditions
Min. Typ. Max. Unit
Vin Input Voltage
fin Input Frequency
Po Output Power Range
Vo Output Voltage Range
Io Output Current Range
Iil Input Current (Low Line)
Iih Input Current (High Line)
Irl Low Line Inrush Current
Irh High Line Inrush Current
Eff Efficiency
Operating Voltage
90
264 VAC
47
63 Hz
Vin=90 to 264VAC
0
12 W
See rating chart V
See rating chart A
Io=Full load, Vin=115VAC
0.25 0.35 A
Io=Full load, Vin=230VAC
0.17 0.22 A
Io=Full load, 25c, Cool start,
Vin=115VAC
Io=Full load, 25c, Cool start,
Vin=230VAC
14 16 A
28 30 A
Io=Full Load, Vin=230VAC
70.8 72 85 %
REG-i Line Regulation
REG-o Load Regulation
Io=Full Load
Vin=230VAC
0.5 1
%
35
%
OVP Over Voltage Protection
112
132 %
OCP Over Current Protection
110
150 %
Ttr Time of Transient Response Io=Full Load to Half Load, Vin=100VAC
4 mS
Thold Hold-Up Time
Io=Full Load, Vin=110VAC
10 16
mS
Ts Start Up Time
Io=Full Load, Vin=100VAC
0.3 1 2
S
Vrn Ripple & Noise(Peak to Peak) Full Load, Vin=90VAC
0.5 1
%
TC Temperature Coefficient
Thermal Shutdown ○1 by
Tjsd Junction Temperature
Controller
All output
The parameter is not subject
to production test-verified
by design/characterization
of integrated controller.
-0.04
-25
0.04 %/c
130 c
○1 As long as faulty conditions have been removed , the adaptor will automatically power up as usual.
Environmental :
Sym.
Parameter
T e s t Conditions
Toper Operating Temperature
Tstg Storage Temperature
Hr Relative Humidity
No-Condensing
MTBF Operating Temperature at 25 C, Calculated per MIL-HDBK-217F
Pd Derate linearly from 100% load at 50 C to 50% load at 70 C
Min. Typ. Max. Unit
0 50 70
C
-40
85
C
5
95 %
0.1
M Hrs
Total Power International, Inc. Toll Free: 877-646-0900(US) www.total-power.com