English
Language : 

TPMPU12A Datasheet, PDF (1/2 Pages) Total Power International – 12W Wall Moumt Switching Power Supplies For Medical Equipment
TPMPU12A SERIES
12W Wall Mount Switching Power Supplies For Medical Equipment
Description:
The 12A series of AC/DC switching mode power supplies
provide 12 watts of continuous output power.
This series is suitable for use in Blood Pressure measurements,
Frequency Therapy Device and Dental Curing light applications.
All models are designed to comply with UL/c-UL(UL60601-1),
and new CE requirements.
Features:
Wide Input Voltage 90 to 264 VAC,47 to 63 Hz
2 Prong Plug-In Mains Connector
Output Voltage Available From 5VDC Thru 27VDC
Optional Output Connector (See appendix)
Single Output
Class II
Over Voltage and Over Load protection.
Over temperature Detection
3 year warranty
Electrical Characteristics:
Safety Approvals:
PS
E
Sym.
Parameter
Test Conditions
Min. Typ. Max. Unit
Vin Input Voltage
Operating Voltage
90
264 VAC
fin Input Frequency
47
63 Hz
Po Output Power Range
Vin=90 to 264VAC
0
12 W
Vo Output Voltage Range
See rating chart V
Io Output Current Range
See rating chart A
Iil Input Current (Low Line) Io=Full load, Vin=115VAC
0.25 0.35 A
Iih Input Current (High Line) Io=Full load, Vin=230VAC
0.17 0.22 A
Irl Low Line Inrush Current
Io=Full load, 25c, Cool start,
Vin=115VAC
14 16 A
Irh High Line Inrush Current
Io=Full load, 25c, Cool start,
Vin=230VAC
28 30 A
Eff Efficiency
Io=Full Load, Vin=230VAC
70.8 72 85 %
REG-i Line Regulation
Io=Full Load
0.5 1
%
REG-o Load Regulation
Vin=230VAC
35
%
OVP Over Voltage Protection
112
132 %
OCP Over Current Protection
110
150 %
Ttr Time of Transient Response Io=Full Load to Half Load, Vin=100VAC
4 mS
Thold Hold-Up Time
Io=Full Load, Vin=110VAC
10 16
mS
Ts Start Up Time
Io=Full Load, Vin=100VAC
0.3 1 2
S
Vrn Ripple & Noise(Peak to Peak) Full Load, Vin=90VAC
0.5 1
%
TC Temperature Coefficient
Thermal Shutdown ○1 by
Tjsd Junction Temperature
Controller
All output
The parameter is not subject
to production test-verified
by design/characterization
of integrated controller.
-0.04
-25
0.04 %/c
130 c
○1 As long as faulty conditions have been removed , the adaptor will automatically power up as usual.
Environmental :
Sym.
Parameter
T e s t Conditions
Min. Typ. Max. Unit
Toper Operating Temperature
Tstg Storage Temperature
Hr Relative Humidity
No-Condensing
MTBF Operating Temperature at 25 C, Calculated per MIL-HDBK-217F
Pd Derate linearly from 100% load at 50 C to 50% load at 70 C
0 50
-40
5
0.1
70
C
85
C
95 %
M Hrs