English
Language : 

TPES22 Datasheet, PDF (1/2 Pages) Total Power International – UNIVERSAL INPUT AC TO DC MEDICAL APPLICATION EXTERNAL ADAPTER SINGLE OUTPUT 22-24 WATTS SWITCHING POWER SUPPLIES TPES22 SERIES
TOTAL POWER INT'L
UNIVERSAL INPUT AC TO DC MEDICAL APPLICATION
EXTERNAL ADAPTER SINGLE OUTPUT 22-24 WATTS
SWITCHING POWER SUPPLIES TPES22 SERIES
FEATURES:
UNIVERSAL AC SOURCES
DESKTOP IEC320 AC RECEPTACLE
MEDICAL STANDARDS EN6061/UL2601
EMI MEET EN55022 / FCC CLASS B
CE MARKING COMPLIANCE
SPECIFICATION
INPUT SPECIFICATION
Input Voltage: Typ. 90-264Vac.
Input Connector: IEC320-3P DT7 AC receptacle.
Input Frequency: 47-63Hz.
Inrush Current: 35A @ 230Vac. typical or less than
30A by adding thermister.
Input Current: Typ. 0.38A @ 115Vac. &
0.15A @ 230Vac.
Dielectric Withstand: Meet IEC601.
EMI: Meet EN 55022 Class B.
Hold-up Time: Typ.16mS @115Vac. or
64mS @230Vac.
Earth Leakage: Less than 0.1 mA.
OUTPUT SPECIFICATION
Output Voltage: See Ratings Chart.
Output Current: See Ratings Chart.
Output Wattage: 22-24W typical.
Output Connector & Cord: Optional.
Line Regulation: Typ. 0.1%.
Load Regulation: Typ. ±1.5-3.0 %.
Noise & Ripple: Typ.1.0% peak to peak.
OVP: Built-in, by latched circuit.
Adjustability: Factory set.
Overload Protection: Typ.115-130% of max. power
of overrange or short circuit, power foldback &
self-recovering.
GENERAL SPECIFICATION
Efficiency: Typ. 80%.
Switching Frequency: Typ. 75K Hz.
Circuit Topology: Fixed Frequency flyback circuit.
Transient Response: Output voltage returns in less than
3mS following a 50% load change.
Safety Standard: IEC601/EN 60601/UL 2601 Class I
for medical applications.
Operating Temperature: 0 to +40°C.
Storage Temperature: -20 to +85°C.
Temperature Coefficient: 0.04% /°C.
Cooling: Free air convection.
Construction: Impact resistant thermo- plastic
enclosure case.
Power Density: 1.75Watts / Cubic Inch.
Desktop construction format.
Notes: (1) All measurements are at nominal input, full load, and +25℃ unless otherwise specified.
(2) Load regulation is measured at 115Vac or 230Vac in percentage to indicate the change in output voltage as the load varied from half load to full load(±%).
(3) The exact obtainable load regulation depends upon the output cord selected and load current. Upper data are for 6 ft. (2 m) cord AWG#18 wires.
In application.
TOTAL POWER INT'L Tel: 877-646-0900 Fax: 978-453-7395
Email: sales@total-power.com http://www.total-power.com
-1-