English
Language : 

TPEMG24 Datasheet, PDF (1/2 Pages) Total Power International – ACCOMMODATE UNIVERSAL AC INPUT
Total Power International, Inc.
UNIVERSAL INPUT AC-DC MEDICAL & ITE APPLICATION
EXTERNAL DESKTOP SWITCHING ADAPTER 20-24 WATTS
GREEN POWER SINGLE OUTPUT TPEMG24 SERIES
FEATURES:
 ACCOMMODATE UNIVERSAL AC INPUT
 MEET MEDICAL STANDARDS IEC60601-1
& ITE STANDARDS IEC60950-1
 EMI MEET EN 55011 & EN55022 /
FCC CLASS B
 MEET ENERGY STAR LEVEL V &
CEC LEVEL IV
SPECIFICATION
INPUT SPECIFICATION
Input Voltage: Typical 90-264Vac.
Input Connector: 3 pole AC inlet IEC320-C14(DT7) /
2 pole AC inlet IEC320-C8(DT8).
Input Frequency: 47-63Hz.
Inrush Current: 6.274Arms at 230Vac.
Input Current: Typical 0.347A at 115Vac/
0.215A at 230Vac.
Dielectric Withstand: Meet IEC60601-1 & IEC60950-1.
EMI: Meet EN55011 & EN55022 / FCC Class B.
Hold-up Time: Typical 16.8mS at 115Vac.
Typical 86.4mS at 230Vac.
Leakage Current: Typical 0.25 mA for Class I
Typical 0.1mA for Class II.
No Load Power: Less than 0.3W.
OUTPUT SPECIFICATION
Output Voltage: See Ratings Chart.
Output Current: See Ratings Chart.
Output Wattage: Typical 20-24Watts.
Output Connector & Cord: Optional.
Line Regulation: Typical 0.1%.
Load Regulation: Typical ±1.8%-±3%.
Noise & Ripple: 1.0% peak to peak.
OVP: Built-in by Auto-recovery.
Adjustability: Factory set.
Over Current Protection (OCP): Installed.
Current limiting can be set precisely as per request.
GENERAL SPECIFICATION
Efficiency: Typical 78%-88% (various with the output
voltage)
Switching Frequency: Typical 55K Hz.
Circuit Topology: Fixed Frequency Flyback circuit.
Transient Response: Output voltage returns in less than
Operating Temperature: 0°C to +40°C.
Storage Temperature: -20 to +85°C.
Cooling: Free air convection.
Construction: Impact resistant thermo-plastic
3mS following a 25% load change.
enclosure case.
Safety Standard: Meet Medical standards UL60601-1/
EN60601-1, and ITE UL60950-1/EN60950-1
Class I for DT7(C14) or Class II for DT8(C8)
Power Density: 1.6-19.1Watts / Cubic inch.
Desktop Format.
NOTE: (1) All measurements are at nominal input, full load, and +25°C unless otherwise specified.
(2) Load regulation is measured at 115Vac or 230Vac in percentage to indicate the change in output voltage as the load varied from half load to full load (%).
(3) The exact obtainable load regulation depends upon the output cord selected and load current.
(4) Due to requests in market and advances in technology, specifications subject to change without notice.
For the details of safety approval, please consult the factory.
Total Power International, Inc. Toll Free: 877-646-0900(US) PH# 978-453-7272 www.total-power.com