English
Language : 

U5200 Datasheet, PDF (3/8 Pages) TE Connectivity Ltd – Industrial Pressure Transducer
U5200
Industrial Pressure Transducer
Ambient Temperature: 25°C (unless otherwise specified)
PARAMETERS
MIN
TYP
-0.5
Accuracy
(RSS of linearity, hysteresis, and repeatability)
-0.25
-0.1
-0.25
-0.75
Isolation, Body to any Lead
100
Dielectric Strength
MAX
0.5
0.25
0.1
0.25
0.75
2
UNITS
%F.S. BFSL
%F.S. BFSL
%F.S. BFSL
%F.S. BFSL
%F.S. BFSL
MΩ
mA
NOTES
≤ 2psi @ 25°C
> 2psi and ≤ 5psi @ 25°C
> 5psi and ≤ 500psi @ 25°C
> 500psi and ≤ 5000psi @ 25°C
> 5000psi @ 25°C
@500VDC
@500VAC, 1min
Pressure Cycles
1.00E+6
0~FS Cycles
Proof Pressure
3X
20k psi
Rated
Burst Pressure
4X
20k psi
Rated
Long Term Stability (1 year)
-0.1
0.1
%F.S.
Total Error Band
Compensated Temperature
-1.25
-1.0
-0.75
-1.25
-20
1.25
%F.S.
1.0
%F.S.
0.75
%F.S.
1.25
%F.S.
+85
°C
≤ 2psi
> 2psi and ≤ 5psi
> 5psi and ≤ 5000psi
> 5000psi
Operating Temperature
-40
+125
°C
Except cable 105°C max
Storage Temperature
-40
+125
°C
Except cable 105°C max
Load Resistance (RL)
Load Resistance (RL)
Current Consumption
RL > 100k
Ω
< (Supply Voltage -9V) / 0.02A
Ω
5
mA
Voltage Output
Current Output
Voltage Output
Rise Time (10% to 90%)
<2ms (Voltage Output); <3ms (Current Output); Without Snubber
Pressure Port Material
316L Stainless Steel; 316L Stainless Steel Snubber
Shock
50g, 11msec Half Sine Shock per MIL-STD-202G, Method 213B, Condition A
Vibration
±20g, MIL-STD-810C, Procedure 514.2, Fig 514.2-2, Curve L
For custom configurations, consult factory.
Notes
Compensated Temperature: The temperature range over which the product will produce an output proportional to pressure within the specified
performance limits.
Operating Temperature: The temperature range over which the product will produce an output proportional to pressure but may not remain
within the specified performance limits.
Storage Temperature: The temperature range over which the product can be stored safely in occasions without pressure applied or power
input and remains rated performance. Beyond this temperature range may cause permanent damage to the product.
All configurations are built with supply voltage reverse and output short-circuit protections.
CE Compliance
EN 55022 Emissions Class A & B
IEC 61000-4-2 Electrostatic Discharge Immunity (8kV contact/15kV air)
IEC 61000-4-3 Radiated, Radio-Frequency Electromagnetic Field Immunity (10V/m, 80M-1GHz)
IEC 61000-4-4 Electrical Fast Transient Immunity (1kV)
IEC 61000-4-5 Surge Immunity (V+ to V-: ±2KV/42Ω; L to Case: ±1KV/12Ω; V- to V0: ±1KV/42Ω)
IEC 61000-4-6 Immunity to Conducted Disturbances Induced by Radio Frequency
Fields (150K~80MHz, 10V level for voltage output models, 3V level for current output model)
IEC 61000-4-9 Pulse Magnetic Field Immunity (100A/m peak)
For all CE compliance tests, max allowed output deviation ±1.5 %F.S.
SENSOR SOLUTIONS ///U5200
08/2016
Page 3