English
Language : 

805M1 Datasheet, PDF (2/3 Pages) TE Connectivity Ltd – TO-5 Series Accelerometer
MODEL 805M1 ACCELEROMETER
PERFORMANCE SPECIFICATIONS
All values are typical at +24°C, 80Hz and 5Vdc excitation unless otherwise stated. Measurement Specialties reserves the right to update
and change these specifications without notice.
Parameters
DYNAMIC
Range (g)
Sensitivity (mV/g)
Frequency Response (Hz)
Frequency Response (Hz)
Natural Frequency (Hz)
Non-Linearity (%FSO)
Transverse Sensitivity (%)
Shock Limit (g)
Residual Noise (g RMS)
±20
100
1-8000
0.4-10000
34000
±1
5
5000
0.001
±50
40
1-8000
0.4-10000
34000
±1
5
5000
0.001
±100
20
1-8000
0.4-12000
34000
±1
5
5000
0.003
±200
10
1-8000
0.4-12000
34000
±1
5
5000
0.003
±500
4
1-8000
0.4-12000
34000
±1
5
5000
0.005
Notes
±10%
±1dB
±3dB
Broadband 1Hz to 10kHz
ELECTRICAL
Bias Voltage (Vdc)
Total Supply Current (mA)
Excitation Voltage (Vdc) 1
Output Impedance (Ω)
Insulation Resistance (MΩ)
Full Scale Output Voltage (V)
Grounding
Exc Voltage / 2
<0.80
3.0 to 5.5
<100
>100
±2
Case grounded to mounting surface
@50Vdc
ENVIRONMENTAL
Temperature Response (%)
Operating Temperature (°C)
Storage Temperature (°C)
±10
-40 to +100
-54 to +125
PHYSICAL
Sensing Element
Case Material
Weight (grams)
Mounting
Piezo-Ceramic
Stainless Steel
5.0
Adhesive or Circuit Board (M3 Stud Mounting for -01 Option)
1 The model 805M1 can be operated with 2.8V excitation but the full-scale range will be limited.
Calibration supplied:
CS-SENS-0100 NIST Traceable Amplitude Calibration at 80Hz
The information in this sheet has been carefully reviewed and is believed to be accurate; however, no responsibility is assumed for inaccuracies. Furthermore, this
information does not convey to the purchaser of such devices any license under the patent rights to the manufacturer. Measurement Specialties, Inc. reserves the
right to make changes without further notice to any product herein. Measurement Specialties, Inc. makes no warranty, representation or guarantee regarding the
suitability of its product for any particular purpose, nor does Measurement Specialties, Inc. assume any liability arising out of the application or use of any product or
circuit and specifically disclaims any and all liability, including without limitation consequential or incidental damages. Typical parameters can and do vary in different
applications. All operating parameters must be validated for each customer application by customer’s technical experts. Measurement Specialties, Inc. does not
convey any license under its patent rights nor the rights of others.
SENSOR SOLUTIONS /// Model 805M1 Rev D
9/2015
Page 2