6EHQ Datasheet, PDF (1/3 Pages) TE Connectivity Ltd – Highest Performance RFI Filters for Medical Equipment | |||
|
Corcom Product Guide
Catalog: 1654001
Issue Date: 06.2011
Highest Performance RFI Filters for Medical Equipment
HQ Series
UL Recognized
CSA Certiï¬ed
VDE Approved
3EHQ3
6EHQ8
1
3EHQ1
HQ Series
⢠Designed to provide the highest available
attenuation of RFI noise in the frequency range
from 10kHz to 30MHz for low leakage current
applications
⢠Size and cost-effective
Ordering Information
6 E HQ 8 M
Suï¬x
M â Metric inserts (style 8 only)
Omit for standard or none
Input / Output Style
1 â .250 [6.3] spade terminals
3 â Wire leads
8 â IEC inlet* (line side)
Wire leads (load side)
HQ Series
Speciï¬cations
Maximum leakage current each Line to Ground:
@ 120 VAC 60 Hz:
2 ïA
@250 VAC 50 Hz:
5 ïA
Hipot rating (one minute):
Line to Ground:
Line to Line:
2250 VDC
1450 VDC
Rated Voltage (max):
250 VAC
Operating Frequency:
50/60 Hz
Rated Current:
3 & 6A
Operating Ambient Temperature Range
(at rated current Ir):
-10°C to +40°C
In an ambient temperature (Ta) higher than +40°C
the
maximum
operating
current (Io)
follows: Io
is calculated as
= Ir ï(85-Ta)/45
Available Part Numbers
3EHQ1
3EHQ3
3EHQ8
3EHQ8M
6EHQ1
6EHQ3
6EHQ8
3EHQ8M
Accessories
GA400: NEMA 5-15P to IEC 60320-1 C-13 line cord
Leakage current designation
E â Low leakage (<0.5 mA)
Current Rating
3 or 6A
*IEC 60320-1 C14 inlet mates with C13 connector
Dimensions are in inches and millimeters unless otherwise speciï¬ed. Values in italics
are metric equivalents. Dimensions are shown for reference purposes only.
Speciï¬cations subject to change.
37
For email, phone or live chat, please go to
te.com/help
corcom.com
|
▷ |