English
Language : 

3MV1 Datasheet, PDF (1/2 Pages) Tyco Electronics – Multi-purpose Medical Filter for Power Line Noise Protection with Limited Leakage Current
Corcom Product Guide
Catalog: 1654001
Issue Date: 06.2011
Multi-purpose Medical Filter for Power Line Noise Protection
MV Series
UL Recognized
CSA Certified
VDE Approved
MV Series
• Multi-purpose medical filter
• Improved Line to Ground performance
• A good solution to emission or immunity problems
• Meets leakage current requirements of UL2601 for
health care equipment
Specifications
Maximum leakage current each Line to Ground:
@ 120 VAC 60 Hz:
.07 mA
@250 VAC 50 Hz:
.13 mA
Hipot rating (one minute):
Line to Ground:
Line to Line:
2250 VDC
1450 VDC
Rated Voltage (max):
250 VAC
Operating Frequency:
50/60 Hz
Rated Current:
3 to 20A
Operating Ambient Temperature Range
(at rated current Ir):
-10°C to +40°C
In an ambient temperature (Ta) higher than +40°C
the
maximum
operating
current (Io)
follows: Io
is calculated as
= Ir (85-Ta)/45
Ordering Information
6 MV 1
Input / Output Style
1 – .250 [6.3] spade terminals
MV Series
Current Rating
3, 6, 10 or 20A
Electrical Schematic
LINE
LOAD
Available Part Numbers
3MV1
10MV1
6MV1
20MV1
54
Dimensions are in inches and millimeters unless otherwise specified. Values in italics
are metric equivalents. Dimensions are shown for reference purposes only.
Specifications subject to change.
For email, phone or live chat, please go to
te.com/help