3EHZ1 Datasheet, PDF (1/3 Pages) Tyco Electronics – High Performance Power Line Filter for Medical Applications
Corcom Product Guide
Catalog: 1654001
Issue Date: 06.2011
High Performance Power Line Filter for Medical Applications
HZ Series
UL Recognized
CSA Certified
1
VDE Approved
HZ Series
• Designed to provide good attenuation to RFI noise
in the frequency range from 10kHz to 30MHz
• Size and cost-effective
• Low leakage current
• New versions up to 30A
Ordering Information
3 E HZ 1
Input / Output Style
1 – .250 [6.3] spade terminals
6 – Threaded bolt
HZ Series
Specifications
Maximum leakage current each Line to Ground:
@ 120 VAC 60 Hz:
2 A
@250 VAC 50 Hz:
5 A
Hipot rating (one minute):
Line to Ground:
Line to Line:
2250 VDC
1450 VDC
Rated Voltage (max):
250 VAC
Operating Frequency:
50/60 Hz
Rated Current:
3 to 30A
Operating Ambient Temperature Range
(at rated current Ir):
-10°C to +40°C
In an ambient temperature (Ta) higher than +40°C
the
maximum
operating
current (Io)
follows: Io
is calculated as
= Ir (85-Ta)/45
Leakage current designation
E – Low leakage (<0.5 mA)
Current Rating
3, 4, 6, 10, 15, 20 or 30A
Case Styles
3EHZ1
Available Part Numbers
3EHZ1
6EHZ1
15EHZ1
30EHZ6
4EHZ1
10EHZ1
20EHZ1
Electrical Schematic
LINE
LOAD
B
D
C
E
A







Dimensions are in inches and millimeters unless otherwise specified. Values in italics
are metric equivalents. Dimensions are shown for reference purposes only.
Specifications subject to change.