English
Language : 

3EHQ Datasheet, PDF (1/3 Pages) TE Connectivity Ltd – Highest Performance RFI Filters for Medical Equipment
Corcom Product Guide
Catalog: 1654001
Issue Date: 06.2011
Highest Performance RFI Filters for Medical Equipment
HQ Series
UL Recognized
CSA Certified
VDE Approved
3EHQ3
6EHQ8
1
3EHQ1
HQ Series
• Designed to provide the highest available
attenuation of RFI noise in the frequency range
from 10kHz to 30MHz for low leakage current
applications
• Size and cost-effective
Ordering Information
6 E HQ 8 M
Suffix
M – Metric inserts (style 8 only)
Omit for standard or none
Input / Output Style
1 – .250 [6.3] spade terminals
3 – Wire leads
8 – IEC inlet* (line side)
Wire leads (load side)
HQ Series
Specifications
Maximum leakage current each Line to Ground:
@ 120 VAC 60 Hz:
2 A
@250 VAC 50 Hz:
5 A
Hipot rating (one minute):
Line to Ground:
Line to Line:
2250 VDC
1450 VDC
Rated Voltage (max):
250 VAC
Operating Frequency:
50/60 Hz
Rated Current:
3 & 6A
Operating Ambient Temperature Range
(at rated current Ir):
-10°C to +40°C
In an ambient temperature (Ta) higher than +40°C
the
maximum
operating
current (Io)
follows: Io
is calculated as
= Ir (85-Ta)/45
Available Part Numbers
3EHQ1
3EHQ3
3EHQ8
3EHQ8M
6EHQ1
6EHQ3
6EHQ8
3EHQ8M
Accessories
GA400: NEMA 5-15P to IEC 60320-1 C-13 line cord
Leakage current designation
E – Low leakage (<0.5 mA)
Current Rating
3 or 6A
*IEC 60320-1 C14 inlet mates with C13 connector
Dimensions are in inches and millimeters unless otherwise specified. Values in italics
are metric equivalents. Dimensions are shown for reference purposes only.
Specifications subject to change.
37
For email, phone or live chat, please go to
te.com/help
corcom.com