10EHT Datasheet, PDF (1/3 Pages) TE Connectivity Ltd – High Performance RFI Power Line Filters for Medical Equipment | |||
|
Corcom Product Guide
Catalog: 1654001
Issue Date: 06.2011
High Performance RFI Power Line Filters for Medical Equipment
HT Series
UL Recognized
CSA Certiï¬ed
VDE Approved
6EHT3
6EHT1
6EHT7
HT Series
Speciï¬cations
⢠Designed to provide signiï¬cant attenuation of RFI
noise in the frequency range from 10kHz to 30MHz
⢠Size and cost-effective
Ordering Information
Maximum leakage current each Line to Ground:
@ 120 VAC 60 Hz:
2 ïA
@250 VAC 50 Hz:
5 ïA
Hipot rating (one minute):
Line to Ground:
Line to Line:
2250 VDC
1450 VDC
Rated Voltage (max):
250 VAC
Operating Frequency:
50/60 Hz
Rated Current:
3 to 15A
Operating Ambient Temperature Range
(at rated current Ir):
-10°C to +40°C
In an ambient temperature (Ta) higher than +40°C
the
maximum
operating
current (Io)
follows: Io
is calculated as
= Ir ï(85-Ta)/45
6 E HT 7 M
Suï¬x
M â Metric inserts (style 7 only)
Omit for standard or none
Input / Output Style
1 â .250 [6.3] spade terminals
3 â Wire leads
6 â Threaded bolt
7 â IEC inlet* (line side)
.250 [6.3] spade terminals
(load side)
HT Series
Available Part Numbers
3EHT1
3EHT3
3EHT7
3EHT7M
6EHT1
6EHT3
6EHT7
6EHT7M
10EHT1
10EHT3
15EHT1
15EHT6
Accessories
GA400: NEMA 5-15P to IEC 60320-1 C-13 line cord
Leakage current designation
E â Low leakage (<0.5 mA)
Current Rating
3, 6, 10 or 15A
*IEC 60320-1 C14 inlet mates with C13 connector
40
Dimensions are in inches and millimeters unless otherwise speciï¬ed. Values in italics
are metric equivalents. Dimensions are shown for reference purposes only.
Speciï¬cations subject to change.
For email, phone or live chat, please go to
te.com/help
corcom.com
|
▷ |