DTM65-C8 Datasheet, PDF (1/2 Pages) TDK Electronics – 40W to 65W Medical Class II AC-DC External Power Supplies | |||
|
DTM65-C8 Series
40W to 65W Medical Class II AC-DC External Power Supplies
Features
ïµ Meets DOE Level VI Efï¬ciency
ïµ 60601-1 & 60601-1-11 Medical Certiï¬cations
ïµ Class II Input
ïµ < 0.21W Off-load Power Draw
Key Market Segments & Applications
Speciï¬cations
Model
Input Voltage range
Inrush Current
Input Current
Power Factor
Hold Up Time (Typ)
Temperature Coefï¬cient
Voltage Accuracy
Adjustment Range
Minimum Load
Total Regulation
Ripple & Noise
Overload & Short Circuit Protection
Overvoltage Protection
Overtemperature Protection
Efï¬ciency
Operating Temperature
Storage Temperature
Humidity (non condensing)
Cooling
Withstand Voltage
Vibration (non operating)
Shock
Safety Agency Certiï¬cations
Ofï¬oad Power Consumption
EMC
IP Rating
Weight (Typ)
Size (WxLxH)
Cable Length
AC Input Connector
Output Connector
VAC
A
A
-
ms
-
%
-
A
-
%
-
V
-
%
-
°C
%RH
-
VAC
-
-
-
W
-
-
g
in (mm)
mm
-
-
Warranty
Years
90 - 264VAC (47 - 63Hz)
<100A at 230VAC input, 25°C ambient cold start
2A at 90VAC, Full Load
Meets EN61000-3-2 Class A
10ms at 115VAC input
±0.05%/°C
±1%
None
None
5V to 15V Models:±5%, >15V models ±3%
1% pk-pk
Continuous - hiccup mode
110 - 150% of nominal (Cycle input power to reset)
Yes (Cycle input power to reset)
>88%, average efï¬ciency (5V model: >83.8%)
0 to +60°C derate linearly to 50% load from 41 to 60°C
-10 to +70°C
10 - 90% RH
Convection
Input to Ground 1.5kVAC, Input to Output 4kVAC
23.52m/s2 (10 - 55Hz: constant sweep 1 min X, Y, Z for 1 hour)
< 196.1 m/s2 (20G)
ANSI/AAMI ES60601-1, EN/IEC60601-1, CSA-C22.2 No.60601 3rd edition (2 x MOPP)
< 0.21W
EN55011B, EN60601-1-2
IP21
240g
2.36 x 4.17 x 1.22â (60 x 106 x 31mm)
1050mm
IEC 320-C8 (2 prong)
Kycon KPP-4P or equivalent
Pins 1 & 2: -Vout, Pins 3 & 4: +Vout
Three Years
1-800-526-2324 ⢠www.us.tdk-lambda.com/lp/
58
Disclaimer: us.tdk-lambda.com/lp/legal.htm
|
▷ |