DTM165-C Datasheet, PDF (1/2 Pages) TDK Electronics – 160W to 165W Medical AC-DC External Power Supplies | |||
|
DTM165-C Series
160W to 165W Medical AC-DC External Power Supplies
Features
ïµ CEC, ErP and EISA(1) level V Compliant Models
ïµ Wide Range AC Input
ïµ Medical Certiï¬cations
ïµ > 87% Average Efï¬ciency
Key Market Segments & Applications
Speciï¬cations
Model
Input Voltage range
Inrush Current
Input Current
Power Factor
Leakage Current (264VAC 60Hz)
Hold Up Time (Typ)
Temperature Coefï¬cient
Voltage Accuracy
Adjustment Range
Minimum Load
Total Regulation
Ripple & Noise
Short Circuit Protection
Overvoltage Protection
Efï¬ciency
Operating Temperature
Storage Temperature
Humidity (non condensing)
Cooling
Withstand Voltage
Vibration (non operating)
Shock
Safety Agency Certiï¬cation
Efï¬ciency Level
Ofï¬oad Power Consumption
Conducted & Radiated EMI
Immunity
Weight (Typ)
Size (WxLxH)
Cable Length
AC Input Connector
Output Connector
Warranty
VAC
A
A
-
uA
ms
%/°C
%
V
A
%
%
-
V
%
°C
°C
%RH
-
VAC
-
-
-
-
W
-
-
g
in
mm
-
-
yrs
DTM165PWxxx-C
90 - 264VAC (47 - 63Hz)
<130A at 230VAC input, 25°C ambient cold start
2.5A at 90VAC, Full Load
Meets EN61000-3-2. >0.9 PF
<300uA
>10ms
±0.05%/°C
±1%
None
None
12V to 19V Models:±5%, >19V models ±3%
1%
<160% - hiccup mode
120 - 150% of nominal (Cycle input power to reset)
>85% average efï¬ciency
0 to +60°C, derate linearly 50% load from 40 to 60°C
-10 to +80°C
0 - 90%RH
Convection
Input to Ground 1.5kVAC, Input to Output 4kVAC (Reinforced)(2 x MOPPs)
(Output is connected to Ground)
23.52m/s2 (10 - 55Hz: constant sweep 1 min X, Y, Z for 1 hour)
< 196.1 m/s2 (20G)
UL60601-1, EN60601-1, IEC60601-1, CE Mark
CEC, ErP, EISA level V
< 0.5W
EN55022, FCC Part 18 Class B
EN55011:2007+A2:2007
790
3.35 x 6.7 x 1.73â
1050mm
IEC 320-C14 (Accepts IEC 320-C13)
Kycon KPPx-4P or equivalent
Pins 1 & 2: -Vout, Pins 3 & 4: +Vout
Two Years
(1) EISA - Energy Independence and Security Act of 2007
1-800-526-2324 ⢠www.us.tdk-lambda.com/lp/
45
Disclaimer: us.tdk-lambda.com/lp/legal.htm
|
▷ |