English
Language : 

CSS500 Datasheet, PDF (1/2 Pages) TDK Electronics – 500W AC-DC Medical or Industrial Power Supplies
CSS500 Series
500W AC-DC Medical or Industrial Power Supplies
‹ 360W Convection rating
‹ High Efficiency
‹ IEC60601-1 or IEC60950-1 certifications
‹ ORing FET & Current Share
Key Market Segments & Applications
Medical
Test and Measurement
Features and Benefits
Feature
‹ High Efficiency
‹ Current Share and ORing FET
‹ Dual input fuses
Benefit
‹ Uses less energy
‹ Use in redundant applications
‹ Easier system compliance
Specifications
ITEMS
Input Voltage range
Inrush Current
Input Current (115/230VAC)
Power Factor Correction
Leakage Current
Hold Up Time (Typ)
Temperature Coefficient
Voltage Accuracy
Adjustment Range
Total Regulation
Ripple & Noise
Overcurrent Protection
Overvoltage Protection
Overtemperature Protection
Remote On/Off
Efficiency
Standby Voltage
Fan Output
Current Share
DC Good & Fan Fail Signals
ORing FET
Operating Temperature
Storage Temperature
Humidity (non condensing)
Cooling
Withstand Voltage
(1)
Isolation Resistance
Vibration (non operating)
Shock
Safety Agency Approvals
Conducted & Radiated EMI
Immunity
MTBF
Weight (Typ)
Size (WxLxH)
Warranty
CSS500
-
90 - 264VAC (47 - 63Hz)
A
<50A maximum at 230VAC input, 25°C ambient cold start
A
6 / 3A
-
Meets EN61000-3-2, >0.9
µA
<300µA 264VAC 63Hz
ms
20ms at 115VAC input, 360W loading
-
±0.04%/°C
%
±1% at 60% load
%
None
-
±5%
%
1% peak to peak (5Vsb also)
-
110 - 150%
V
110 - 130% of nominal (Cycle input power or use remote on/off to reset)
-
Yes (Cycle input power or use remote on/off to reset)
-
Unit on: Floating or high 3.5 - 5.25V, Unit off: Low or <0 - 0.5V
%
87% to 92%, model & input dependant
-
5V 1A (5V 0.25A when convection cooled or when inhibit is activated)
-
12V 1A
-
Single wire, up to 4 units can be shared within 10% accuracy at full load
-
Both Low on Fail
-
Yes, for redundant operation
°C
Convection (U channel): 0 to +70°C, derate linearly to 35% load from 40 to 70°C
Forced air (or internal fan): 0 to +70°C, derate linearly to 50% load from 50 to 70°C
°C
-10 to +85°C
%
10 - 95% RH
-
Convection or forced air (30CFM or 3.1m/s). Internal fan is temperature controlled
- Input to Ground 1.5kVAC, Input to Output 4kVAC(1), Output to Ground 500VAC for 1 min.
-
>20M at 25°C & 70%RH, Output to Ground 500VDC
-
19.6m/s2 (10~55Hz:2G Constant, X,Y,Z 60min each.)
-
< 196.1 m/s2 (20G)
-
Medical Version: UL60601-1, IEC60601-1, Industrial Version: UL60950-1, EN60601-1, CE Mark
-
Medical Version: EN55011-B, FCC Class B, Industrial: EN55022, FCC Class B
-
Medical Version: EN60601-1-2, Industrial: EN55024
-
>68,695 (MIL-217F-HDBK)
g
340g (U channel)
in
U Channel: 8 x 4.7 x 1.51", End Fan (/S) 9 x 4.7 x 1.63", Top Fan (/T) 8 x 4.7 x 2.85"
yr
Two Years
Note (1) Industrial version (I suffix): 3kVAC
3055 Del Sol Blvd • San Diego, CA 92154 • 1-800-526-2324
29