CSS500 Datasheet, PDF (1/2 Pages) TDK Electronics – 500W AC-DC Medical or Industrial Power Supplies | |||
|
CSS500 Series
500W AC-DC Medical or Industrial Power Supplies
 360W Convection rating
 High Efficiency
 IEC60601-1 or IEC60950-1 certifications
 ORing FET & Current Share
Key Market Segments & Applications
Medical
Test and Measurement
Features and Benefits
Feature
 High Efficiency
 Current Share and ORing FET
 Dual input fuses
Benefit
 Uses less energy
 Use in redundant applications
 Easier system compliance
Specifications
ITEMS
Input Voltage range
Inrush Current
Input Current (115/230VAC)
Power Factor Correction
Leakage Current
Hold Up Time (Typ)
Temperature Coefficient
Voltage Accuracy
Adjustment Range
Total Regulation
Ripple & Noise
Overcurrent Protection
Overvoltage Protection
Overtemperature Protection
Remote On/Off
Efficiency
Standby Voltage
Fan Output
Current Share
DC Good & Fan Fail Signals
ORing FET
Operating Temperature
Storage Temperature
Humidity (non condensing)
Cooling
Withstand Voltage
(1)
Isolation Resistance
Vibration (non operating)
Shock
Safety Agency Approvals
Conducted & Radiated EMI
Immunity
MTBF
Weight (Typ)
Size (WxLxH)
Warranty
CSS500
-
90 - 264VAC (47 - 63Hz)
A
<50A maximum at 230VAC input, 25°C ambient cold start
A
6 / 3A
-
Meets EN61000-3-2, >0.9
µA
<300µA 264VAC 63Hz
ms
20ms at 115VAC input, 360W loading
-
±0.04%/°C
%
±1% at 60% load
%
None
-
±5%
%
1% peak to peak (5Vsb also)
-
110 - 150%
V
110 - 130% of nominal (Cycle input power or use remote on/off to reset)
-
Yes (Cycle input power or use remote on/off to reset)
-
Unit on: Floating or high 3.5 - 5.25V, Unit off: Low or <0 - 0.5V
%
87% to 92%, model & input dependant
-
5V 1A (5V 0.25A when convection cooled or when inhibit is activated)
-
12V 1A
-
Single wire, up to 4 units can be shared within 10% accuracy at full load
-
Both Low on Fail
-
Yes, for redundant operation
°C
Convection (U channel): 0 to +70°C, derate linearly to 35% load from 40 to 70°C
Forced air (or internal fan): 0 to +70°C, derate linearly to 50% load from 50 to 70°C
°C
-10 to +85°C
%
10 - 95% RH
-
Convection or forced air (30CFM or 3.1m/s). Internal fan is temperature controlled
- Input to Ground 1.5kVAC, Input to Output 4kVAC(1), Output to Ground 500VAC for 1 min.
-
>20M at 25°C & 70%RH, Output to Ground 500VDC
-
19.6m/s2 (10~55Hz:2G Constant, X,Y,Z 60min each.)
-
< 196.1 m/s2 (20G)
-
Medical Version: UL60601-1, IEC60601-1, Industrial Version: UL60950-1, EN60601-1, CE Mark
-
Medical Version: EN55011-B, FCC Class B, Industrial: EN55022, FCC Class B
-
Medical Version: EN60601-1-2, Industrial: EN55024
-
>68,695 (MIL-217F-HDBK)
g
340g (U channel)
in
U Channel: 8 x 4.7 x 1.51", End Fan (/S) 9 x 4.7 x 1.63", Top Fan (/T) 8 x 4.7 x 2.85"
yr
Two Years
Note (1) Industrial version (I suffix): 3kVAC
3055 Del Sol Blvd ⢠San Diego, CA 92154 ⢠1-800-526-2324