English
Language : 

AQ0300MU12EACFD Datasheet, PDF (2/13 Pages) SynQor Worldwide Headquarters – Medical Grade AC/DC Power Supply With PFC
Technical Specification
AC Input: 85-264 VRMS
DC Output: 12/24/36/48V Semi-reg.
Power: 300 W Series
Grade: Medical
ACuQor 300W Series ELECTRICAL CHARACTERISTICS
All specifications typical with TA = 25 °C, unless otherwise specified.
MAIN OUTPUT SPECIFICATIONS
GENERAL SPECIFICATIONS
Output power (continuous) 85-132/170-264 Vrms
(5 s transient)
85-132/170-264 Vrms
132-170 Vrms
Nominal DC output
voltage (at 250W)
(Semi-regulated)
12 Vout
24 Vout
36 Vout
48 Vout
Efficiency (see figs. 3 - 10) 12 Vout, 115 Vrms, 300 W
48 Vout, 115 Vrms, 300 W
12 Vout, 230 Vrms, 300 W
48 Vout, 230 Vrms, 300 W
Hold-up time (to -20%)
12 Vout
24 / 36 / 48 Vout
Maximum load capacitance 12 Vout
24 Vout
36 Vout
48 Vout
Output ripple voltage
Switching frequency (20 MHz BW)
Twice line frequency (at 300W)
300 W
400 W
See Figure 10
12.4 V
25 V
37.5 V
50 V
89% typ.
90% typ.
90% typ.
91% typ.
16 ms @ 300 W
20 ms @ 300 W
16,000 µF
8,000 µF
4,000 µF
2,000 µF
0.5% p-p
5.0% p-p
Fundamental ripple freq. Input
Output
500 kHz
250 kHz
Audible noise
Fan speed varies with temp.
39 dBA @ 1 m max.
Weight (EA \ EC)
(SC \ RC)
(UC \ TC)
343 g (12.1 oz) \ 446 g (15.7 oz)
846 g (29.8 oz) \ 879 g (31 oz)
1224 g (43.2 oz) \ 1298 g (45.8 oz)
MTBF
MIL-217
Demonstrated
343.6 kHours
TBD kHours
ISOLATION SPECIFICATIONS
Isolation voltage
Input to output
Input to ground
Output to ground (BF & CF )
Output to ground (CFD)
4000 Vrms
1500 Vrms
1500 Vrms
5000 Vpulse
Insulation resistance
Output to ground
10 MΩ min.
Leakage currents
See Note 2
ENVIRONMENTAL CHARACTERISTICS
Thermal performance
Operating ambient (see Figure 9)
Non-operating ambient
0 ºC to +70 ºC
-40 ºC to +85 ºC
Turn-on delay
Transient response
Iout steps from 50-75%
At 0.2 A/µs
Overvoltage protection
Cyclic restart
Short circuit protection
Cyclic operation
Total regulation
Over line, load and temperature
Auxillary Output
Always on (See Note 1)
Thermal protection
Automatic recovery
REMOTE_ENABLE
Input Low Voltage
Input High Voltage
INPUT SPECIFICATIONS
AC input voltage
Universal range
Input frequency
Input current
Power factor
115 Vrms @ 400 W
230 Vrms @ 400 W
2 s max.
3% typ / 6% max.
dev.
100 ms recovery
110-120%
115% rated Iout
±6.0%
5 V @ 50 mA
+125 ºC (PCB Temp)
0.45 V (max)
4.15 V (min)
85-264 Vrms
47-63 Hz
3 Arms
1.5 Arms
>0.98
Relative humidity
Non-condensing
Altitude
Operating
Non-operating
Random vibration
5-500 Hz
Shock
Half-sine, 10 ms, 3 axes
EMC CHARACTERISTICS
Conducted emissions
EN55011 and EN55022, FCC
part15
Line frequency harmonics EN61000-3-2
Voltage fluctuations
EN61000-3-3
ESD air
EN61000-4-2
ESD contact
EN61000-4-2
Radiated immunity
EN61000-4-3
Fast transients
EN61000-4-4
Line surge immunity
EN61000-4-5
Conducted immunity
EN61000-4-6
Power freq. mag. field
EN61000-4-8
5-95% RH
10,000 ft max.
30,000 ft max.
0.03 g2/Hz
20 g peak
Level B
Class A
Clause 5b
Level 3
Level 3
Level 3
Level 3
Level 3
Level 3
3 A/m
Input surge current
Internal input fuses
264 Vrms (cold start)
Both AC lines
40 A max.
6.3 A
Voltage dip immunity
EN61000-4-11
Perf Criteria A, A, B
<5% UT 10 ms,
70% UT 500 ms,
40% UT 100 ms
SAFETY AGENCY CERTIFICATIONS (pending)
All certification marks appear on individual unit labels.
UL60601-1 and UL60950
IEC/EN 60601-1 and IEC/EN 60950
CE Marked
cUL CSA Standards
Meets NFPA 99 2005 300 µA earth leakage
NOTES:
1. Derate 1 mA per °C above 50 °C ambient temperature.
2. Leakage currents see page 4.
Product # AQ0300MUxx Series Phone 1-888-567-9596
www.synqor.com
Doc.# 005-005044 Rev. A
09/09/08
Page 2