English
Language : 

ME30_16 Datasheet, PDF (2/5 Pages) SL Power Electronics – 20 to 30W Single Output External Power
ME30
20 to 30W Single Output External Power
Medical Series
Model Selection (continued)
Model
Output Output Ripple &
Line
Load
Overvoltage
Output
Input
Number
Volts Current Power
Noise¹
Regulation Regulation Trip Range Connector Configuration
ME30A0503Q01 5.0V 4.00A
20W 75mV pk-pk
±1%
±5%
5.75V-7.75V
ME30A0903Q01
ME30A1203Q01
ME30A1503Q01
ME30A1803Q01
ME30A2403Q01
9.0V
12.0V
15.0V
18.0V
24.0V
3.00A
2.50A
2.00A
1.67A
1.33A
27W
30W
30W
30W
30W
90mV pk-pk
120mV pk-pk
150mV pk-pk
180mV pk-pk
240mV pk-pk
±1%
±1%
±1%
±1%
±1%
±5%
±5%
±5%
±5%
±5%
11.7V-16.2V
14.4V-16.8V
18.0V-21.0V
21.6V-25.2V
28.8V-33.6V
2.5 x 5.5 x
9.5mm
Straight Barrel
Type, center
positive
Class II
Desktop,
IEC60320 C18
Receptacle
ME30A4803Q01 48.0V 0.63A
30W 480mV pk-pk
±1%
±5%
55.0V-60.0V
ME30A0503B01 5.0V 4.00A
20W 75mV pk-pk
±1%
±5%
5.75V-7.75V
ME30A0903B01
ME30A1203B01
ME30A1503B01
ME30A1803B01
ME30A2403B01
9.0V
12.0V
15.0V
18.0V
24.0V
3.00A
2.50A
2.00A
1.67A
1.33A
27W
30W
30W
30W
30W
90mV pk-pk
120mV pk-pk
150mV pk-pk
180mV pk-pk
240mV pk-pk
±1%
±1%
±1%
±1%
±1%
±5%
±5%
±5%
±5%
±5%
11.7V-16.2V
14.4V-16.8V
18.0V-21.0V
21.6V-25.2V
28.8V-33.6V
2.5 x 5.5 x
9.5mm
Straight Barrel
Type, center
positive
Class II Wall-
Plug,
Interchangeabl
e Blades (North
American Blade
included)²
ME30A4803B01 48.0V 0.63A
30W 480mV pk-pk
±1%
±5%
55.0V-60.0V
ME30A0503C01 5.0V 4.00A
20W 75mV pk-pk
±1%
±5%
5.75V-7.75V
ME30A0903C01
ME30A1203C01
ME30A1503C01
ME30A1803C01
ME30A2403C01
9.0V
12.0V
15.0V
18.0V
24.0V
3.00A
2.50A
2.00A
1.67A
1.33A
27W
30W
30W
30W
30W
90mV pk-pk
120mV pk-pk
150mV pk-pk
180mV pk-pk
240mV pk-pk
±1%
±1%
±1%
±1%
±1%
±5%
±5%
±5%
±5%
±5%
11.7V-16.2V
14.4V-16.8V
18.0V-21.0V
21.6V-25.2V
28.8V-33.6V
2.5 x 5.5 x
9.5mm
Straight Barrel
Type, center
positive
Class II Wall-
Plug, Fixed
North American
Blades³
ME30A4803C01 48.0V 0.63A
30W 480mV pk-pk
±1%
±5%
55.0V-60.0V
Notes: 1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors. For 5V and
6V models, values listed are typical, 100mV pk-pk maximum with 0.1μF ceramic and 47μF low ESR capacitors used at measurement point.
2. Order blade kit KT-1027K for other blades (EU. UK, Australia)
3. For EU fixed blades, replace “C” in the model number with “M”, for UK blades, replace “C” with “G”, for Australia blades, replace “C” with “H”.
4. All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
5. For Input Class I models: For AC GND connected to output common (-), insert a “B” in the part number where the “A” is located (ME30B1203F01).
General Specifications
AC Input 100-240Vac, ±10%, 47-63Hz, 1∅
Turn On Time Less than 700mS @115Vac, full load
Input Current 115Vac: 1.2A, 230Vac:0.6A
Inrush Current
Input Fuses
Earth Leakage
Current
Efficiency
264Vac, cold start: will not exceed 40A
F1, F2: 2.0A, 250Vac fuses (line & neutral
lines) provided on all models
Input-GND: <500μA@264Vac, 60Hz, NC
Output-GND: <4mA@264Vac, 60Hz, NC
>87%, typical
Output Power
No Load Input
Power
20 to 30W continuous – See models chart for
specific voltage model ratings.
<0.1W per DoE Efficiency Level VI
Requirements
Ripple and Noise See models chart on pg 1.
Output Voltage
Transient
Response
Regulation
Drop Test
See models chart on pg 1.
500µs response time for return to within 0.5%
of final value for any 50% load step over the
range of 5% to 100% of rated load, ∆i/∆t<
0.2A/µs. Max. voltage deviation is +/-3.5%.
See models chart on pg 1.
1.4m from table top to wooden platform, 6
faces.
Hold-up Time
Overtemperature
Protection
Overload Protection
Short Circuit
Protection
Overvoltage
Protection
Isolation
Safety Standards
Operating
Temperature
Temperature
Derating
20mS min., at full Load, 100Vac input
Will shutdown upon an overtemperature
condition, auto-recovery.
130 to 180% of rating, Hiccup Mode
Hiccup Mode, auto recovery.
Hiccup mode, see models chart for trip
ranges.
Input-Output: 2 MOPP
Input-Ground: 1 MOPP
Output-Ground: 1 MOPP
EN/IEC/UL60601-1-1, 3rd edition
-20°C to +50°C
Start Up at -40°C, full load, (warmup period before
all parameters are within published specifications).
Derate output power above 40°C to TBD at
50°C
Storage Temperature -40°C to +85°C
Altitude
Relative Humidity
Operating: to 5000m.
Non-operating: -500 to 40,000 ft.
5% to 95%, non-condensing
www.slpower.com
ME30 Rev 1.9 25-APR-16
2