English
Language : 

ME20 Datasheet, PDF (2/4 Pages) SL Power Electronics – 20W Single Output External Power Medical Series
ME20
20W Single Output External Power
Medical Series
Model Selection (continued)
Model
Output Output Ripple &
Line
Load
Output
Input
Number
Volts Current Power
Noise¹
Regulation Regulation
Connector
Configuration
ME20A0503Q01 5.0V 3.00A
15W 75mV pk-pk
±1%
±5%
ME20A0703Q01 7.5V 2.00A
15W 75mV pk-pk
±1%
±5%
ME20A0903Q01
ME20A1203Q01
ME20A1503Q01
ME20A1803Q01
9.0V
12.0V
15.0V
18.0V
2.00A
1.50A
1.20A
1.00A
18W
18W
18W
18W
90mV pk-pk
120mV pk-pk
150mV pk-pk
180mV pk-pk
±1%
±1%
±1%
±1%
±5%
±5%
±5%
±5%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Desktop,
IEC60320 C18
Receptacle
ME20A2403Q01 24.0V 0.83A
20W 240mV pk-pk
±1%
±5%
ME20A4803Q01 48.0V 0.42A
20W 480mV pk-pk
±1%
±5%
ME20A0503B01 5.0V 3.00A
15W 75mV pk-pk
±1%
±5%
ME20A0703B01 7.5V 2.00A
15W 75mV pk-pk
±1%
±5%
ME20A0903B01
ME20A1203B01
ME20A1503B01
ME20A1803B01
9.0V
12.0V
15.0V
18.0V
2.00A
1.50A
1.20A
1.00A
18W
18W
18W
18W
90mV pk-pk
120mV pk-pk
150mV pk-pk
180mV pk-pk
±1%
±1%
±1%
±1%
±5%
±5%
±5%
±5%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Wall-Plug,
Interchangeable
Blades (North American
Blade included)²
ME20A2403B01 24.0V 0.83A
20W 240mV pk-pk
±1%
±5%
ME20A4803B01 48.0V 0.42A
20W 480mV pk-pk
±1%
±5%
ME20A0503C01 5.0V 3.00A
15W 75mV pk-pk
±1%
±5%
ME20A0703C01 7.5V 2.00A
15W 75mV pk-pk
±1%
±5%
ME20A0903C01
ME20A1203C01
ME20A1503C01
ME20A1803C01
9.0V
12.0V
15.0V
18.0V
2.00A
1.50A
1.20A
1.00A
18W
18W
18W
18W
90mV pk-pk
120mV pk-pk
150mV pk-pk
180mV pk-pk
±1%
±1%
±1%
±1%
±5%
±5%
±5%
±5%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Wall-Plug,
Fixed North American
Blades³
ME20A2403C01 24.0V 0.83A
20W 240mV pk-pk
±1%
±5%
ME20A4803C01 48.0V 0.42A
20W 480mV pk-pk
±1%
±5%
Notes: 1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors. For 5V and
6V models, values listed are typical, 100mV pk-pk maximum with 0.1μF ceramic and 47μF low ESR capacitors used at measurement point.
2. Order blade kit KT-1027K for other blades (EU. UK, Australia)
3. For EU fixed blades, replace “C” in the model number with “M”, for UK blades, replace “C” with “G”, for Australia blades, replace “C” with “H”.
4. For Input Class I models: For AC GND connected to output common (-), insert a “B” in the part number where the “A” is located (ME20B0503F01).
5. All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
General Specifications
AC Input 100-240Vac, ±10%, 47-63Hz, 1∅
Turn On Time Less than 700mS @115Vac, full load
Input Current 115Vac: 0.5A, 230Vac: 0.25A
Inrush Current 264Vac, cold start: will not exceed 40A
Input Fuses
F1, F2: 3.15A, 250Vac fuses (line & neutral
lines) provided on all models
Earth Leakage Input-GND: <500μA@264Vac, 60Hz, NC
Current Output-GND: <4mA@264Vac, 60Hz, NC
Efficiency >87%, typical
Output Power
15 to 20W continuous – See models chart for
specific voltage model ratings.
No Load Input <0.1W per DoE Efficiency Level VI
Power Requirements
Ripple and Noise See models chart on pg 1.
Hold-up Time 20mS min., at full Load, 100Vac input
Overtemperature Will shutdown upon an overtemperature
Protection condition, auto-recovery.
Overload Protection 130 to 180% of rating, Hiccup Mode
Short Circuit
Protection
Hiccup Mode, auto recovery.
Overvoltage
Protection
130 to 150% of output voltage, hiccup mode
Isolation
Safety Standards
Operating
Temperature
Input-Output: 2 MOPP
Input-Ground: 1 MOPP
Output-Ground: 1500Vac
ANSI/AAMI ES60601-1:2005/(R)2012, CSA
CAN/CSA-C22.2 NO, 60601-1:14; IEC60601-
1:2005+CORR.2:2007+AM:2012; EN 60601-
1:2006/A11:2011, EN60601-1: 2006/A12:2014,
EN60601-1:2006/A1:2013,
-20°C to +50°C
Start Up at -40°C, full load, (warmup period before
all parameters are within published specifications).
www.slpower.com
ME20 Rev 2.4 31-JAN-16
2