ME10 Datasheet, PDF (2/4 Pages) SL Power Electronics – 10W-12W Single Output External Power Medical Series
ME10
10W-12W Single Output External Power
Medical Series
Model Selection (continued)
Model
Output Output Ripple &
Line
Load
Output
Input
Number
Volts Current Power
Noise¹
Regulation Regulation
Connector
Configuration
ME10A0503C01 5.0V
2.0A
10W 75mV pk-pk
±1%
±5%
ME10A0603C01 5.9V
1.6A
10W 75mV pk-pk
±1%
±5%
2.5 x 5.5 x 9.5mm
Class II Wall-Plug,
ME10A0703C01 7.5V
ME10A1203C01 12.0V
1.3A
1.0A
10W
12W
75mV pk-pk
120mV pk-pk
±1%
±1%
±5%
±5%
Straight Barrel Type,
center positive
Fixed North American
Blades³
ME10A2403C01 24.0V 0.5A
12W 240mV pk-pk
±1%
±5%
Notes: 1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors. For 5V and
6V models, values listed are typical, 100mV pk-pk maximum with 0.1μF ceramic and 47μF low ESR capacitors used at measurement point.
2. Order blade kit KT-1027K for other blades (EU. UK, Australia)
3. For EU fixed blades, replace “C” in the model number with “M”, for UK blades, replace “C” with “G”, for Australia blades, replace “C” with “H”.
4. For Input Class I models: For AC GND connected to output common (-), insert a “B” in the part number where the “A” is located (ME10B0503F01).
5. All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
General Specifications
AC Input 100-240Vac, ±10%, 47-63Hz, 1∅
Turn On Time Less than 700mS @115Vac, full load
Input Current 115Vac: 0.45A, 230Vac: 0.28A
Hold-up Time 20mS min., at full Load, 100Vac input
Inrush Current 264Vac, cold start: will not exceed 40A
Overtemperature Will shutdown upon an overtemperature
Protection condition, auto-recovery.
Input Fuses
F1, F2: 3.15A, 250Vac fuses (line & neutral
lines) provided on all models
Overload Protection 130 to 180% of rating, Hiccup Mode
Earth Leakage Input-GND: <500μA@264Vac, 60Hz, NC
Current Output-GND: <4mA@264Vac, 60Hz, NC
Efficiency >87%, typical
Output Power
10 to 12W continuous – See models chart for
specific voltage model ratings.
No Load Input <0.1W per DoE Efficiency Level VI
Power Requirements
Ripple and Noise See models chart on pg 1.
Output Voltage See models chart on pg 1.
Transient
Response
500µs response time for return to within 0.5%
of final value for any 50% load step over the
range of 5% to 100% of rated load, ∆i/∆t<
0.2A/µs. Max. voltage deviation is +/-3.5%.
Regulation See models chart on pg 1.
Short Circuit
Protection
Hiccup Mode, auto recovery.
Overvoltage
Protection
130 to 150% of output voltage, hiccup mode
Isolation
Safety Standards
Input-Output: 2 MOPP
Input-Ground: 1 MOPP
Output-Ground: 1500Vac
EN/IEC/UL60601-1, 3rd edition and
EN/IEC/UL60601-1-11:2010 for Home
Healthcare
Operating
Temperature
-20°C to +50°C
Start Up at -40°C, full load, (warmup period before
all parameters are within published specifications).
Temperature Derate output power above 40°C to TBD at
Derating 50°C
Storage Temperature -40°C to +85°C
Altitude
Operating: to 5000m.
Non-operating: -500 to 40,000 ft.
Drop Test
1.4m from table top to wooden platform, 6
faces.
Relative Humidity 5% to 95%, non-condensing
Vibration
E-Cap Life
www.slpower.com
Operating: 0.003g/Hz, 1.5grms overall, 3
axes, 10 min/axis, 1-500Hz.
Non-Oper.: random waveform, 3 minutes per
axis, 3 axes and Sine waveform, Vib.
frequency/acceleration: 10-500Hz/1g, sweep
rate of 1 octave / minutes, Vibration time of 10
sweeps / axes, 3 axes
>10 year life based on calculations at
115Vac/60Hz & 230Vac/50Hz, ambient 25°C
at 24 hrs per day, 365 days/year, 6 power up
cycles per day.
ME10 Rev 2.5
31-JAN-16
Shock
Operating: Half-sine, 20gpk, 10mS, 3 axes,
6 shocks total
Non-Operating: Half-sine waveform, impact
acceleration of 100G, Pulse duration of 6
mS, Number of shocks: 3 for each of the
three axis
MTBF
>1,000,000 hours, full load, 110 & 220Vac
input, 25°C amb., per Telcordia 332 Issue 6,
Stress Method.