English
Language : 

MU425S Datasheet, PDF (1/3 Pages) SL Power Electronics – 425W Single Output
MU425S
425W Single Output
Medical Power Supply
Features
 3.3” x 6.2” x 1.62” Package
 Up to 425W of Air-Cooled Power, 300W Convection
 Universal Input 90-264Vac Input Range
 5V at 2A Standby Output
 12V Fan Output
 Inhibit, Power Fail, DC OK Signals, Remote Sense
 Approved to CSA/EN/IEC/UL60601-1, 3rd Edition
2 MOPP Isolation
 Meets EN61000-4-2, EN61000-4-3 and EN61000-4-6
Requirements for Home Healthcare Applications
 Efficiency 88% typical
 3-Year Warranty
 RoHS Compliant
Description
The MU425 family is designed to power the latest professional healthcare devices. All models meet 2 MOPP input-to-output
isolation. Models are available with main output voltages of 12V, 18V, 24V and 48V.
Model Selection
Model
Output Current
Ripple & Noise³
Total
Efficiency4
OVP
Number
Volts
w/200LFM air¹ Convection²
(mV pk-pk)
Regulation
(Main Output)
Threshold
12V
32.2A
22.0A
120mV
±3%
88%
14.0 ± 1.1V
MU425S12E
5Vsb
2.0A
2.0A
12V Fan6
1.0A
0.5A
100mV
360mV
±5%
±10%5
5.5V – 8.0V
N/A
18V
21.45A
14.6A
180mV
±3%
88%
21.0 ± 2.0V
MU425S18E
5Vsb
2.0A
2.0A
12V Fan6
1.0A
0.5A
100mV
360mV
±5%
±10%5
5.5V – 8.0V
N/A
24V
16.8A
11.9A
240mV
±3%
90%
28.0 ± 2.5V
MU425S24E
5Vsb
2.0A
2.0A
12V Fan6
1.0A
0.5A
100mV
360mV
±5%
±10%5
5.5V – 8.0V
N/A
48V
8.4A
5.9A
480mV
±3%
90%
58.0 ± 2.0V
MU425S48E
5Vsb
2.0A
2.0A
12V Fan6
1.0A
0.5A
100mV
360mV
±5%
±10%5
5.5V – 8.0V
N/A
Notes: 1. Total power with 200 lfm of forced air cooling is 425W (408W for 18V, 385W for 12V) including 12V/1A for Fan output and 5V/2A standby.
2. Total convection power is 300W (280W for 12V model). Total power includes 5Vsb and Fan outputs.
3. Measured at 25°C ambient with noise probe directly at end of 6” twisted pair terminated with 0.1μF ceramic and 10μF low ESR capacitors. Values will be higher at
ambient temperatures below 0°C.
4. Efficiency values listed are typical and are measured at 115Vac input, full load output current, at an ambient temperature of 25°C.
5. Fan output regulation is with 1A min load on main output.
6. Fan Output: If the load on the output is other than a fan, a short circuit condition on the output can only be remedied by removing both the cause of the short circuit and
the load. This will allow the output to resume normal operation.
General Specifications
100-240Vac, ±10%, 47-63Hz, 1∅
AC Input
120–300Vdc (external fuse required for DC input)
Input Current 115Vac: 5.2A, 230Vac: 2.5A
264Vac, cold start: will not exceed 40Arms
Inrush Current
within ½ cycle. I²T = 25A²/Sec maximum
Input Fuses F1, F2: 6.3A, 250Vac
Efficiency See chart above.
Turn On Time
(Main Output)
Hold-up Time
Overtemperature
Protection
Overload
Protection
Switching
Frequency
Main output: <1 sec. max @115Vac, rise time 30mS max.
5Vsb turn-on time is 500mS max., rise time 50mS max.
Output Voltage rise is monotonic.
Main Output: >20ms for 300W @ 120Vac/60 Hz, >16ms
for 367W (90% of 408W) @ 120v/60Hz.
5Vsb Output: >500mS
Sensing transformer temperature, 135°C (55°C ambient
temperature at full load), auto recovery.
130 to 170% of rating, Hiccup Mode, auto-recovery.
75kHz, typical
www.slpower.com
MU425S REV 3.1 14-APR-16
1