English
Language : 

MINT1500B Datasheet, PDF (1/6 Pages) SL Power Electronics – 500W Single Output LED Series
MINT1500B
500W Single Output
Medical Series
Features
3.3” x 7.17” x 1.5 Package, Ideal for 1U
Applications
Up to 500W w/air, 350W convection cooled
Universal Input 90 to 276Vac
(85Vac input 400Watts with 200 LFM)
1500Vac isolation (1MOPP) between output to
ground for BF applications
(Applies to all outputs and signals)
Active current share
Class I Input
Standby and Fan Output Voltages
Inhibit, Power Fail, Output OK Signals
Approved to IEC60601-1, 3rd Edition
2 x MOPP Isolation
93% Efficiency @ 230V Input, 92% @ 115V
Input
Optional Cover & With fan option
3 mounting orientations
3 Year Warranty
Low inrush Current
Description
A superior performance 500 Watt AC to DC power supply designed for Medical applications. Feature rich and highly
efficient, MINT1500 product family with active current share for redundant applications can easily fit in 1U or 2U chassis
and provides 350 Watts without cover for convection, or 500 Watts with moving air. Input & output monitoring alarms plus
12V/0.8A fan output and 5V standby voltage are standard features of the MINT1500B family. All models are CE marked to
low voltage directive and approved to IEC60601-1 3rd edition.
Model Selection
Model
Number1
Volts
MINT1500B2414E01 24V
Output Current
w/200LFM air Convection
20.8A
14.6A
Fan
Output
12Vdc/0.8A
5V Standby
Output
5V@200mA
Ripple &
Noise2
1%
Total
Regulation
±2%
OVP
Threshold
27.6 ± 1.0V
MINT1500B4814E01 48V
10.4A
7.3A
12Vdc/0.8A 5V@200mA
1%
±2%
55.5 ± 2.0V
Notes:
1) To order the required options such as fan or cover, please see below designation:
MINT1500BWW14XYZ
X: E or L(input connector)
Y: 0( U Channel, Standard), C (Cover),T(Top fan option), E(End fan Option)
Z: 1(Standard product), 2 through 9 is reserved for customization and will be assigned by SL Power.
2) Measured with noise probe directly across output terminals with 0.1µF ceramic and 10µF low ESR capacitors. For main output load of less than
5%, total noise& ripple will increase to 2%. For ambient temperature of -10°C, minimum of 3 minutes warm up is required to meet the published spec.
www.slpower.com
MINT1500B REV 11.0, 4-22-15