English
Language : 

MINT1400A Datasheet, PDF (1/3 Pages) SL Power Electronics – 400W Single Output Series
MINT1400A
400W Single Output Series
Medical – Universal Input Power Supply
•Compact 3.94” x 7” x 1.58”, 1U compatible
•9.5 W/in³
•300W Convection Rating
•Efficiency up to 91%
•12Vdc to 48Vdc outputs, up to 400W
•Remote ON/OFF by PS-ON Signal
•Safety approved to Medical and ITE Standards
• EN/IEC/UL60601-1 and CSA 601.1
•CSA/EN/IEC/UL60950-1, 2nd Edition
•Compliant to EN55011/22 Conducted Noise,
Class B
•RoHS Compliant – Category 6
2 Year Warranty
Specifications
All Specifications are typical at nominal input, full load at 25°C unless otherwise stated.
AC Input
80-270VAC, 47-63 Hz single phase. (See Note 1) Turn On Time
500mSec maximum
Input Current Input
120VAC: 4.2A, 230VAC: 2.1A Hold-up Time
20 ms at 400W
Inrush Current Input
Active circuit, 37A max Overvoltage Protection
See chart
Input Fuses
F1, F2: 6.3A Overload Protection
Cycling type, 470W typical
Power Factor
Efficiency
0.99 typical, 0.96 minimum Overtemperature Protection
Provided on all models, auto-
Typical 88% @ 120VAC, 91% @ 240VAC
recovery. See page 2 for cooling information
Remote Sense
Output Power
Ripple and Noise
Compensates for up to 500mV
drop on output leads
400W with fan cooling. See derating chart for other
conditions. Peak: 450W for 1 min., 10% duty cycle
1% pk-pk max., 20MHz BW, differential mode
Switching Frequency
Safety Standards
Isolation
Output Voltages
See chart. Voltages adjustable +/- 5% Earth Leakage Current
Total Regulation
See chart Operating Temperature
Minimum Load
Not required Storage Temperature
Transient Response
500µs to 1%, 0.2A/µs, 3.5% Deviation Relative Humidity
Shock and Vibration
Shock: MIL-STD-810E, 516.4 part IV Altitude
Vibration: 5Hz – 500Hz, 2.5 Grms
Model Number Key
MINT 1 400 A 12 10 L 01
Output: 170kHz ± 8kHz;
PFC: 90kHz, ± 8kHz; Bias: 100kHz, ± 6 kHz
Medical: EN/IEC/UL 60601-1, CSA 601.1
ITE: CSA/EN/IEC/UL 60950-1, 2nd Edition
Input-Output: 4000VAC,
Input-Ground: 1500VAC, Output-Ground
100µA at 120VAC, 60Hz Normal.
400µA at 264Vac, 60Hz SFC.
0° to 70°C, refer to derating
curve for ratings >40°C
-40 to +85°C
5% to 95%, non-condensing
Operating: -500 to 10,000 ft.
Non-Operating: -500 to 40,000 ft.
01 = Standard Model, 02 and higher indicates a m odified m odel.
"L" = Input Connector - EBY P/N 5002-03-N-12
"10" = Output Connections - Bus Bars w ith #10 screw s
Output Voltage: "12" = 12V output, "24" = 24V output, etc.
Model Configuration: "A" = Original Configuration
Output Pow er (Watts) - "400" = 400W
# of Outputs
Product Fam ily: "M" = Medical, "I" = Internal, "NT" = New Technology
______________________________________________________________________________________________________________________________ 1 of 3
SL Power Electronics Corp • 6050 King Drive Ventura, CA 93003 • Phone 805.486.4565 • Fax 858.712.2040 • Email: info@slpower.com • www.slpower.com